Functional checks
LOGIQ e – Basic Service Manual
4-23
5461614-100 English Rev. 6
B Mode Softmenu Key
5
Activate Auto
Optimize
Optimize the image based upon a specified
region of interest or anatomy. Press the Center
Button in the Gain Dial to toggle the ATO/ACE
On and Off.
7
Adjust Time Gain
Compensation
(TGC)
Amplifies the returning signals to correct for
the attenuation caused by tissues at increasing
depth. TGC slide pots spaced proportional to
the depth. Move the slide pots to the left/right
to decrease/increase TGC. A TGC curve
appears on the display.
8
Adjust Scan Area
Widen or narrow the size of the sector angle to
maximize the image’s region of interest (ROI).
Press Scan Area and move the Trackball to
narrow/widen the angle.
9
Adjust Zoom
Changes the location of the focal point(s). A
triangular focus marker indicates the depth of
the focal point.
10
Reverse
Toggles the left/right orientation of the scan
image.
Table 4-1: B Mode Operator Panel Controls
Step
Task
Expected Result(s)
Comments
Table 4-2: B Mode Softmenu Key
Step
Task
Expected Result(s)
Comments
1
Adjust Rejection
Selects a level below which echoes will not be
amplified (an echo must have a certain
minimum amplitude before it will be
processed).
2
Activate Colorize
Enables gray scale image colorization. To
deactivate, reselect a Gray Map.
3
Adjust Edge
Enhance
Edge Enhance brings out subtle tissue
differences and boundaries by enhancing the
gray scale differences corresponding to the
edges of structures. Adjustments to M Mode's
edge enhancement affects the M Mode only.
4
Activate Gray Map
Determines how the echo intensity levels
received are presented as shades of gray.
5
Adjust Frequency
Multi Frequency mode lets you downshift to the
probe's next lower frequency or shift up to a
higher frequency.
Summary of Contents for 5483518
Page 5: ...LOGIQ e Basic Service Manual i 3 5461614 100 English Rev 6 ...
Page 6: ...i 4 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 7: ...LOGIQ e Basic Service Manual i 5 5461614 100 English Rev 6 ...
Page 8: ...i 6 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 9: ...LOGIQ e Basic Service Manual i 7 5461614 100 English Rev 6 ...
Page 10: ...i 8 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 11: ...LOGIQ e Basic Service Manual i 9 5461614 100 English Rev 6 ...
Page 12: ...i 10 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 24: ...i 22 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 60: ...Introduction 1 36 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 80: ...Site Preparations 2 20 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 134: ...System Setup 3 54 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 208: ...Components and Functions Theory 5 14 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 262: ...Replacement Procedures 8 26 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 328: ...Care and Maintenance 10 44 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 384: ...Docking Cart Setup 11 56 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 426: ...Index 4 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 ...
Page 427: ...GE ...