background image

DU-AG-0018-25 Rev J 4/19/17 GCX Corp Page 7 of 8  

WARNING 
•  Do not attempt to adjust height without instrument being mounted on Arm. 
 
•  Use caution when disengaging the Height Locking Lever. The total mounted load may have 

changed due to the addition or removal of equipment (other than primary instrument) while 
the Height Locking Lever was engaged. A change in the total load can cause a sudden 
downward or upward movement of the Arm when the Height Locking Lever is disengaged.

 

Adjusting Height of the Mounted Instrument 

 
 
 
 
 

 
 
 
 
1.  Grasp mounted instrument on sides (or handle) and move to desired height. When Arm is counterbalanced correctly 

(page 4), it will maintain position when adjusted to a new height.  

Note: 

Disengagement/re-engagement of the Height 

Locking Lever may be necessary during vertical adjustment of mounted instrument (see 

WARNING

 above). 

Engagement of the Height Locking Lever is not required during normal operation of the counterbalanced instrument. 

 

 
 

 
 

 
 

 

 
 
 
 
 
 
 
 

 
 
 
 
 

Positioning VHM Arm in Channel 

1.  Lock the Arm position using the Height Locking Lever. 

 

2.  Using the 1/8'' hex wrench provided, loosen four (4) set screws in Slide. 

 

3.  Relieve the weight of the mounted instrument by lifting against the underside of 

the Arm, near the instrument.  Simultaneously, push up or pull down the 
opposite end of the Arm immediately in front of the Slide. 

 

4.  Tighten four (4) set screws. 
 

Use caution when disengaging the Height Locking 
Lever.  The  total  mounted  load may  have  changed 
due to the addition or removal of equipment (other 
than primary instrument) while the Height Locking 
Lever was engaged. A change in the total load can 
cause  a  sudden  downward  or  upward  movement 
of  the  Arm  when  the  Height  Locking  Lever  is 
disengaged. 

 

Instrument Must Be Mounted

 

 

Summary of Contents for VHM

Page 1: ...ht of the mounted device The Arm housing should never be disassembled by non GCX personnel Additionally the mounted device should only be removed with the Arm in its highest vertical position and with...

Page 2: ...Philips supplied Table Mount to Mounting Adapter with three 3 M6 x 12mm FHMS see below Mounting VHM Arm in Channel 1 Thread the Height Locking Lever into the mounting hole in the right side of the Ar...

Page 3: ...itor in accordance with the Philips Table Mount instruction sheet CAUTION Ensure the device mounting locking mechanism is properly engaged at all times Fixed Mount 1 Fasten Spacer to Mounting Adapter...

Page 4: ...or FMS 1 Align three 3 mounting holes in Lower Post Disc with three 3 threaded holes in Swivel Cup below left Using the 5 32 hex wrench provided fasten Post to Swivel Cup with three 3 10 32 x 3 8 SHCS...

Page 5: ...ough the Adjuster Cover below center Lock Arm when Counterbalance Adjuster Bolt is accessible 3 With the 1 2 13mm socket wrench on the Adjuster Bolt carefully unlock and support the arm Turn the Adjus...

Page 6: ...cup to gain access to the Swivel Tension Nut Tilting the Mounted Instrument and Adjusting Tilt Tension Installation Note Shown without mounted device for detail purposes 1 To tilt the instrument gras...

Page 7: ...e necessary during vertical adjustment of mounted instrument see WARNING above Engagement of the Height Locking Lever is not required during normal operation of the counterbalanced instrument Position...

Page 8: ...sive material to clean the mounting assembly 4 Damage caused by the use of unapproved substances or processes will not be covered by warranty We recommend testing of any cleaning solution on a small a...

Reviews: