Installing the Wall Channel Cover 

Install the Wall Channel Cover in accordance with the 

Channel Cover Installation Guide (DU-UT-0001-20).

 

 
 
 
VHM Arm Operation & Adjustment 
 
Regardless of model number, ensure the device mounting locking mechanism is properly engaged at all times. 

 

– Use caution. 
 
Swiveling the Mounted Instrument and Adjusting Swivel Tension 

1.  To swivel mounted instrument, simply push or pull instrument. 
 
2.  To adjust swivel tension, tighten or loosen the Swivel Tension Nut using the 1/2'' 

[13 mm] socket wrench provided.  

Installation Note:

 It may be necessary to remove an existing down post from 

the swivel cup to gain access to the Swivel Tension Nut.  

 
 

 
 

Tilting the Mounted Instrument and Adjusting Tilt Tension 

Installation Note: 

Shown without mounted device for detail purposes. 

1.  To tilt the instrument, grasp the top and bottom of the instrument and tilt it to the desired angle. 

 

2.  To adjust tilt tension, equally tighten or loosen the two (2) Tilt Tension screws using the 5/32'' hex wrench provided. 

 

 
 
 
 

 

 

 

 

 

 

 

 

Pivoting the Arm and Adjusting Pivot Tension 

1.  To pivot Arm at Channel, simply push on side of Arm. 

 

2.  To adjust pivot tension, remove top and bottom plastic 

bolt caps from Pivot Bolt. Using the 1/2" [13mm] socket wrench 
provided and a 1/2'' [13mm] box or open-end wrench, tighten or 
loosen Pivot Bolt and Hex Nut to desired tension. Reinstall bolt 
caps back over each end of Pivot Bolt. 

 
 

 
 
 
 
 
 

 

Swivel Tension Nut  

Tilt Tension Screws (2)  

1/2" [13 mm] Socket Wrench 

1/2" [13 mm] Box-end Wrench 

Summary of Contents for VHM

Page 1: ...ht of the mounted device The Arm housing should never be disassembled by non GCX personnel Additionally the mounted device should only be removed with the Arm in its highest vertical position and with...

Page 2: ...Philips supplied Table Mount to Mounting Adapter with three 3 M6 x 12mm FHMS see below Mounting VHM Arm in Channel 1 Thread the Height Locking Lever into the mounting hole in the right side of the Ar...

Page 3: ...itor in accordance with the Philips Table Mount instruction sheet CAUTION Ensure the device mounting locking mechanism is properly engaged at all times Fixed Mount 1 Fasten Spacer to Mounting Adapter...

Page 4: ...or FMS 1 Align three 3 mounting holes in Lower Post Disc with three 3 threaded holes in Swivel Cup below left Using the 5 32 hex wrench provided fasten Post to Swivel Cup with three 3 10 32 x 3 8 SHCS...

Page 5: ...ough the Adjuster Cover below center Lock Arm when Counterbalance Adjuster Bolt is accessible 3 With the 1 2 13mm socket wrench on the Adjuster Bolt carefully unlock and support the arm Turn the Adjus...

Page 6: ...cup to gain access to the Swivel Tension Nut Tilting the Mounted Instrument and Adjusting Tilt Tension Installation Note Shown without mounted device for detail purposes 1 To tilt the instrument gras...

Page 7: ...e necessary during vertical adjustment of mounted instrument see WARNING above Engagement of the Height Locking Lever is not required during normal operation of the counterbalanced instrument Position...

Page 8: ...sive material to clean the mounting assembly 4 Damage caused by the use of unapproved substances or processes will not be covered by warranty We recommend testing of any cleaning solution on a small a...

Reviews: