Version 1.5

 

For hygienic reasons, the sponges must always be washed following termination of the treatment session 

(for this, see chapter 6.2.2)

 
 

 
 

5 Troubleshooting 

5.1 Error 

Messages 

The following failures will be indicated optically by 

NEOSERV 114

.  

 

• 

Overtemperature LED flashing 

 

1.  If overtemperature is very high (90 - 100 °C), the red LED will start flashing and the pump will be 

switched off immediately. 

2.  As soon as the pump has cooled down, the alarm lamp will go off automatically. Now the unit is 

again ready for operation. 

3.  If overtemperature occurs even though the unit has been operated only for a short time (less than one 

hour), at any rate contact your service representative to have the device checked. 

 

• 

Water level indicator flashing 

 

The water level within the water separator is too high. 

See chapter 3.4

 for draining the water separator. 

 
 
 

5.2 Other 

Troubles 

Trouble

 

Potential Cause

 

No or insufficient sucking action 

– 

Foreign particles have accumulated inside the suction 
electrode(s) 

– 

Suction electrodes draw in secondary air 

– 

Leakage in tubing (electrode lines) 

– 

Controls (front panel) in lowest position 

– 

Internal fault (Notify Technical Service!) 

Red LED

 

(overtemperature) 

flashing 

– 

Pump has run hot 

(see chapter 5.1)

 

– 

Check suction electrodes for potential foreign particles 
(clogging) 

– 

Replace suction electrodes if necessary 

– 

Internal fault (Notify Technical Service!) 

Green LED (POWER ON lamp) 
will not come on (or is not alight) 

– 

Unit not switched on; mains cable defective or not plugged in 

– 

Internal fault (Notify Technical Service!) 

– 

Power failure in the treatment room (main fuse) 

No stimulation effect generated 
(suction electrodes) 

– 

Electrotherapy unit switched off 

– 

Electrotherapy unit not connected to the 

NEOSERV 114

 

– 

Connecting cable, electrode line or electrode defective 

– 

Electrotherapy unit defective (Notify technical Service!) 

– 

Internal fault (Notify Technical Service!) 

– 

Stimulation current too low 

– 

Sponges dry 

 

 
In any other case of trouble or failure contact immediately gbo Medizintechnik AG

 

or one of the service 

agencies authorized by gbo Medizintechnik AG. 

Summary of Contents for NEOSERV 114

Page 1: ...gbo Medizintechnik AG 2004 Version 1 5 Suction Wave Massage Unit NEOSERV 114 NEOSERV 114 H User Manual...

Page 2: ...omissions All rights reserved No part of this manual may be reproduced in any form or by any means electronic mechanical or otherwise without the prior written permission of the gbo Medizintechnik AG...

Page 3: ...Medizintechnik AG 2004 Version 1 5 3 Warnings and safety precautions Warning Warnings which have to be observed by all means Caution Observe the instructions for use Note Information that will facili...

Page 4: ...nnector Unit 11 4 OPERATING THE UNIT 12 4 1 Switching on the Unit 12 4 2 Select Channels 12 4 3 Vacuum Setting 12 4 4 Pulses Setting 12 4 5 Applying the Suction Electrodes 13 4 6 Terminating the Treat...

Page 5: ...nt selected is applied to the patient by means of the suction electrodes The NEOSERV 114 works as suction wave massage unit Liquids dirt or bacteria which are sucked up by the suction electrodes are c...

Page 6: ...6 NEOSERV 114 gbo Medizintechnik AG 2004 Version 1 5 1 2 Short Instructions Figure 1 NEOSERV 114 Figure 2 NEOSERV 114 in combination with DUODYNATOR 119...

Page 7: ...LED of the Standby key will be flashing 3 Connect all needed electrodes to the appropriate connectors by means of the electrode hoses red channel I green channel II Note If you use only 2 suction elec...

Page 8: ...t session remove the suction electrodes from the skin and switch off all units or put them into standby mode 10 Clean the sponges on a daily basis 11 Observe the Operating Instructions relating to any...

Page 9: ...ovided to the unit s connector socket power supply unit One end of each electrode hose is connected to the suction connectors obey colors Then connect the suction electrode to the other end of the fle...

Page 10: ...rheating the pump will be automatically switched off and the LED will be flashing 3 4 Water Level If the water level within the water separator is too high the red water level indicator LED will be fl...

Page 11: ...irtight 3 6 Vacuum and Pulse Regulator By means of the Vacuum regulator a continuously variable setting of the vacuum will be achieved The Pulse regulator is for setting the number of suction pulses p...

Page 12: ...pressure points otherwise the suction connection is not airtight 4 3 Vacuum Setting The Vacuum regulator is used for selecting the desired vacuum pressure This vacuum pressure level can be set anywher...

Page 13: ...in 5 Apply the electrodes to the skin and hold them in place until the sucking action sets in Note Each suction electrode can easily be taken off by raising the electrode e g for changing their the po...

Page 14: ...ent sucking action Foreign particles have accumulated inside the suction electrode s Suction electrodes draw in secondary air Leakage in tubing electrode lines Controls front panel in lowest position...

Page 15: ...user 6 1 Safety Controls Irrespective of the legal rules it is recommended to have the device checked by the manufacturer or by a service agency authorized by him at 12 months intervals The check shal...

Page 16: ...ponges may impair the sponges conductivity To improve restore conductivity we recommend using a saline solution Remove the metal part simply by drawing off from the electrode Remove the rubber cover f...

Page 17: ...hone or place them in a distance of 3 m from the device Cardiac pacemakers can extremely be disturbed In these cases the therapy should be only carried out under continuous pulse and ECG control If th...

Page 18: ...nformity sign Caution Observe the instructions for use Application part ungrounded protection degree Type BF This product complies with WEEE Directive 2002 96 EG waste electrical and electronic equipm...

Page 19: ...Relative humidity 30 75 Transport and storage Temperature range 5 C 50 C Relative humidity 90 none condensing gbo Medizintechnik AG reserves the right to modify the design and specification without p...

Page 20: ...as scheduled by the manufacturer is 10 years NEOSERV 114 114 H is an electronic device For its disposal the according regulations for electronic devices have to be observed Incidentals have to be disp...

Page 21: ...nity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete...

Page 22: ...80 MHz to 800 MHz d 2 3 P for 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation di...

Page 23: ...NEOSERV 114 gbo Medizintechnik AG 2004 Version 1 5 23...

Page 24: ...24 NEOSERV 114 gbo Medizintechnik AG 2004 Version 1 5 Art No 014 7 0054...

Reviews: