For hygienic reasons, the sponges must always be washed following termination of the treatment session
.
5 Troubleshooting
5.1 Error
Messages
The following failures will be indicated optically by
NEOSERV 114
.
•
Overtemperature LED flashing
1. If overtemperature is very high (90 - 100 °C), the red LED will start flashing and the pump will be
switched off immediately.
2. As soon as the pump has cooled down, the alarm lamp will go off automatically. Now the unit is
again ready for operation.
3. If overtemperature occurs even though the unit has been operated only for a short time (less than one
hour), at any rate contact your service representative to have the device checked.
•
Water level indicator flashing
The water level within the water separator is too high.
for draining the water separator.
5.2 Other
Troubles
Trouble
Potential Cause
No or insufficient sucking action
–
Foreign particles have accumulated inside the suction
electrode(s)
–
Suction electrodes draw in secondary air
–
Leakage in tubing (electrode lines)
–
Controls (front panel) in lowest position
–
Internal fault (Notify Technical Service!)
Red LED
(overtemperature)
flashing
–
Pump has run hot
–
Check suction electrodes for potential foreign particles
(clogging)
–
Replace suction electrodes if necessary
–
Internal fault (Notify Technical Service!)
Green LED (POWER ON lamp)
will not come on (or is not alight)
–
Unit not switched on; mains cable defective or not plugged in
–
Internal fault (Notify Technical Service!)
–
Power failure in the treatment room (main fuse)
No stimulation effect generated
(suction electrodes)
–
Electrotherapy unit switched off
–
Electrotherapy unit not connected to the
NEOSERV 114
–
Connecting cable, electrode line or electrode defective
–
Electrotherapy unit defective (Notify technical Service!)
–
Internal fault (Notify Technical Service!)
–
Stimulation current too low
–
Sponges dry
In any other case of trouble or failure contact immediately gbo Medizintechnik AG
or one of the service
agencies authorized by gbo Medizintechnik AG.