PREMIE™ Blue S108.250 |
User Manual
Premie Blue Features
| 15
5. Perform IO exercises.
6. When the training session is
completed, open the drainage
tube 2 adjustable clamp and
drain the luid. If needed, use
the dispensing syringe full with
air to push remaining luid out
the system.
If blood mix is used, lush the system with 2-3 syringes full of clean water.
Holes in the bone(s) that are made by the IO needle can be sealed
with “Superglue”.
Replacing the Tibia Bones
Items needed:
•
Tibia bone skin cover
•
Replacement tibia bone
•
Drainage container
1.
Drain all the luids from the system.
2. Remove the syringe connector
from the illing tube and clamp
from the draining tube.
3. Remove the tibia cover.