5. Using OMNI® 2

5.1 Main Screen

OMNI® 2 opens at the Favorites 

Page upon start up.

Changing a Vital Sign

Vital signs can be quickly changed from the Favorites Page or from the 

“Vitals” page. Follow the steps below to change one of the parameters:

1. 

Tap on a vital sign. 

2.  Adjust the vital sign by 

utilizing the slider or the +/- 

buttons.

3.  Select “APPLY NOW”. 

Refer to the OMNI® 2 User Guide 

for more information on changing vital signs.

Summary of Contents for PREMIE Blue S108.250

Page 1: ...ntiic All Rights Reserved www gaumard com PREMIE Blue S108 250 is an interactive educational system developed to assist a certiied instructor It is not a substitute for a comprehensive understanding o...

Page 2: ......

Page 3: ...onditions 2 Storage Conditions 2 Procedures 2 Cleaning 2 2 Getting Started 3 2 1 Overview 3 Airway 3 Appearance 3 Artiicial Respiration 3 Circulation 3 Simulator 3 Venous Access 4 2 2 Terminology 4 Fa...

Page 4: ...5 Umbilical Catheterization 17 Directions for Use 17 Heelstick 18 5 Using OMNI 2 19 5 1 Main Screen 19 Changing a Vital Sign 19 5 2 Hypoxia 20 5 3 CPR 21 Feedback Graphics 21 Actions 22 Calibration 2...

Page 5: ...rkmanship Electronic transmission is warranted to be free from defects at the moment of transmission Your sole and exclusive remedy and Gaumard s sole liability is to replace the defective media or to...

Page 6: ...PREMIE Blue S108 250 User Manual 6 Introduction...

Page 7: ...arts are available from Gaumard Scientiic or from your Distributor IV Arm Only use Gaumard s provided simulated blood Any other simulated blood containing sugar or any additive may cause blockage and...

Page 8: ...ocedures Do not attempt to intubate without lubricating the airway adjunct with mineral oil lubricant Failure to lubricate the device will make intubation very diicult and is likely to result in damag...

Page 9: ...aneuver Appearance Color responds to hypoxic events and interventions healthy mild cyanosis severe cyanosis Articulated head neck jaw arms and legs Artiicial Respiration Bilateral lung expansion with...

Page 10: ...Venous Access Patent umbilicus Intraosseous access at tibia IV training arm 2 2 Terminology Facilitator The person conducting the simulation an instructor or lab staf member Provider A person partici...

Page 11: ...ted on Premie Blue s left side and then connect the power supply to the wall outlet Always operate Premie Blue with the power supply connected OMNI 2 Connection OMNI 2 controls Premie Blue s responses...

Page 12: ...ecting the simulator features 4 Follow the on screen tutorial for a brief overview of the features of OMNI 2 5 Move onto the next steps in the Tutorial by selecting GOT IT 6 Exit the Tutorial at any t...

Page 13: ...torial OMNI 2 will automatically proceed to the Favorites page and establish a connection to Premie Blue WARNING Do not connect Premie Blue or OMNI 2 to a computer LAN network or unathorized diagnosti...

Page 14: ...t sized mask having a thick seal Intubation Premie Blue airway can be intubated orally and nasally using a 2 5 mm ID Endotracheal tube An LMA can be used to ventilate Premie Blue orally WARNING Nasoga...

Page 15: ...as IV access in the lower right arm that can be used for bolus or intravenous infusion as well as for draining luids The lower arm contains a vein groove which is itted with a soft tube closely simula...

Page 16: ...ustable clamp Drainage container not included in kit 2 Fill the dispensing syringe with luid 3 Connect the syringe to the illing tube 4 Allow luid to low through the system and into the drainage conta...

Page 17: ...lature system Use a 25 27 gauge needle set IV Maintenance Replacing the arm skin 1 Gently pull the IV arm skin 2 Add a small amount of lubri cant to the arm 3 Slide the replacement skin onto the arm R...

Page 18: ...black electrical con nector is visible 4 Unfasten the connector by pressing on the clip 5 The male electrical connector resides in the upper arm 6 Locate the replacement item 7 Take the female connec...

Page 19: ...simulator 11 Place the screw into the other side of the elbow joint and fasten until the connection is secure 4 3 Intraosseous Access Setting up an intraosseous access line is an invasive procedure th...

Page 20: ...ure in the bone using inlet and drain valves Directions for Use Tube 1 illing Tube 2 drainage tube 1 Place the end of tube 2 into a drainage container and open the adjustable clamp 2 Fill the dispensi...

Page 21: ...sh remaining luid out the system If blood mix is used lush the system with 2 3 syringes full of clean water Holes in the bone s that are made by the IO needle can be sealed with Superglue Replacing th...

Page 22: ...ue Features 4 Pull out the tibia bone out of the leg 5 Insert the new tibia The illing and drainage tubes are irst followed by the bone 6 Re attach tibia cover 7 Connect the syringe connector to the i...

Page 23: ...determine whether they meet or exceed conventional neonatal CPR standards If acceptable the OMNI 2 tablet causes the skin to turn to a more healthy color if inadequate or non existent OMNI 2 causes t...

Page 24: ...uid to low through the system and into the drainage container 6 Once the luid is seen draining close the adjustable clamp For catheterization use an umbilical catheter WARNING Lubricate the distal tip...

Page 25: ...ing a Vital Sign Vital signs can be quickly changed from the Favorites Page or from the Vitals page Follow the steps below to change one of the parameters 1 Tap on a vital sign 2 Adjust the vital sign...

Page 26: ...stan dards To start a hypoxia session follow the steps below 1 Select the initial state of the simulator Choose from none peripheral or central cyano sis Premie Blue will display full cyanosis or par...

Page 27: ...R by monitoring cadence and depth of cardiac compressions and airway ventilations in real time WARNING Do not perform mouth to mouth ventilation since the simulator airway can be diicult to clean afte...

Page 28: ...e make sure to calibrate the com pressions Perform the following to calibrate the compressions 1 Tap the gear on the up per right of the screen 2 Select Calibration Buttons Function Begin and end a CP...

Page 29: ...8 250 User Manual Using OMNI 2 23 3 Select Compression once the calibration window appears 4 Adjust the Target compression depth with the and tap START 5 Follow the prompts to calibrate the target com...

Page 30: ...E Blue S108 250 User Manual 24 Using OMNI 2 6 Select SAVE to apply the changes 7 Tap Start to begin a CPR session For more information about the CPR monitoring and reporting refer to the OMNI 2 User G...

Page 31: ...rchase with OMNI 2 Follow the steps below in order to activate this feature 1 Connect the provided router to the wall 2 Tap the gear on the top right of the screen 3 Select WIFI Connection Setup 4 Con...

Page 32: ...itor 8 Follow steps 9 10 if the activation code was not inputted 9 Verify that the serial num ber corresponds to the manikin 10 Input a valid activation code The activation code is case sensitive Note...

Page 33: ...itors software 15 Tap the V on the top left and then select Comm Setup 16 Verify that the IP address in the Wireless Communication Setup matches the Omni IP and Omni Port noted before 17 Select Connec...

Page 34: ...nd re placement parts and their prices Item Type Code Artiicial Blood Concentrate Consumable S108 812 Lubricant Consumable S108 974 Arm Skin Replacement S108 813 L M D Lower Intravaneous Right Trainin...

Page 35: ...the right and click Forget the network Manikin is not paired Disconnect the simulator from the power supply turn of bluetooth on OMNI 2 tablet Plug the simulator back in and turn on bluetooth Power s...

Page 36: ...subsequent purchasers or users of the product acquire the product as is and this limited warranty does not apply This limited warranty applies only to the products manufactured and produced by Gaumar...

Page 37: ...e Seatbelt Susie Krash Kids Premie UNI Omni Omni 2 SmartSkin are trademarks of Gaumard Scientiic Company Always dispose of this product and its components in compliance with local laws and regulations...

Reviews: