www.gastron.com
06
_
07
3. Specification
GTD-1000Tx(M) toxic gas detector has been developed to detect gas leaked from industrial sites and various toxic gases
generated from factories, gas storages, and manufacturing processes that produce or use toxic gases and to prevent
accidents in advance.
GTD-1000Tx(M) toxic gas detector is installed in areas with gas leak hazards and continuously monitors gas leak. It
provides DC 4-20 mA standard output signal and RS-485 network signal. Also, DC 4~20 mA standard output enables
transmission distance of the output signal between detector and receiver at maximum 2,500 m (when using cables with
CVVS or CVVSB 1.5 sq or higher). RS-485 network signal can be transmitted up to 1,000 m (when using a designated
cable for RS-485). GTD-1000Tx(M) toxic gas detect must be used at a height below 1,000 m above sea level.
Body of GTD-1000Tx(M) is made of Aluminum alloy and the gas sensor module is made of stainless steel. It consists of
a complete explosion-proof enclosure (Ex d IIC T6). This product can be installed in all areas prone to toxic gas leak and
explosion hazards.
Internal configuration consists of 1 terminal PCB board that outputs measurements externally (DC 4-20 mA, RS-485).
External configuration consists of detector part that monitors gas leak and cable inlets.
It can perform calibration using detachable display (LCD or OLED), therefore maintenance is convenient.
2. Configuration
1. Overview
GTD-1000Tx(M)
Instruction Manual
[Figure 1. GTD-1000Tx(M) Overview]
Gas
Sensor Module
(GSA860EX)
Main Board
(Transmitter)
+24V DC
4-20mA Analog
Signal Output
Transmitter enclosure
Safety Power Supply Unit
(24VDC)
Safety Controller
(PLC or DCS)
GTD-1000TX(M) Gas Detector
Hazardous Area (Filed)
Safe Area
제어
Signal
External Interface
Power Line
Legend
※ Note1. Refer to the measured gas list for measured gases and their ranges. Contact us for special gas.
3.1. Basic Specifications
3.2. Mechanical Specifications
ITEMS
SPECIFICATION
Measuring Type
Diffusion
Measuring Method
- Electro-Chemical Cell
- Heated-semiconductor Cell
- Photoionization detector(PID)
Detectible Gas
Toxic Gas (Note1)
Measuring Range
Capable to display 000.0 ~ 10000 (Note 1)
Accuracy
≤ ±3% / Full Range
Zero Drift
≤2% / Full Range
Response Time
Depends on Sensor Module.
Refer to Sensor Specification or Contact in case for Special Gas.
Approvals Classification)
KCs: Ex d llC T6,T5,T4, IP65
Basic Interface
Analog 4-20mA current interface
Rs-485 Interface
RS-485 MODBUS (Option)
Option
RS-485 Board
GTL-100 Warning Light
Rain Cover
Warranty
Transmitter
2Year
Sensor
1Year
ITEMS
SPECIFICATION
Eexplosion Proof type
Explosion-proof enclosure
Dimension
136(W) × 176(H) × 95 (D) mm
Weight including Sensor
App. 1.5kg
Mounting type
Wall mount
Mounting Holes
Ø 7 ±0.1
Cable inlet
3/4" PF ( 1/2" or 3/4" NPT )
Body material
Transmitter
aluminum alloy
Sensor
Stainless Steel (STS316)