GTD-1000Tx(M)

Instruction Manual

7. System Mode

7.2.2. Span Calibration

CALIBRA.

MODE

- Contact "↑" or "↓" key to select "Calibration Mode".
- Contact "FUNC" key when "CALIBRA. MODE" is displayed to enter Calibration Mode.

CALIBRA.

[SPAN]

- Contact "↑" or "↓" key to achieve [SPAN] then contact "FUNC" key to enter Span Calibration mode.

SPAN CAL

[ NO]

- Contact "↑" or "↓" key to achieve [YES] then contact "FUNC" key to perform Span Calibration.

SPAN GAS

[     50]

- Using a calibration tool, inject the standard gas to the sensor at a flow rate of 500 mL/min for 90 sec. 

After the measurement is stabilized, contact "FUNC" key to enter the next mode.

- When Hold function in Maintenance Mode is on, the current maximum SPAN gas value is held and 

displayed in the first row.

- The current measurement is displayed in the second row.

[      50]
[      50]

SPAN SET

[     50]

- It is a mode to set standard gas value. When there is no "Fail" message, contact "↑" or "↓" key to 

set a value.

- When the inject standard gas value is abnormal, it displays Fail message. Fail message is as follows. 

Fail message and span set value are displayed alternatively.

① "LOW FAIL" occurs when the difference between injected gas value and zero calibration value is 

        below 1%.

② " HIGH FAIL" occurs when injected gas value is above 95% of total ADC input range.

- "RING FAIL" occurs when injected gas value runs in the current set span value, it exceeds 95% of ADC 

input range. The above FAIL messages are automatically released if the status is normal after adjusting 

SPAN value.

SPAN SET
LOW FAIL

SPAN SET

HIGH FAIL

SPAN >>

[SUCESS]

- After completing standard gas value setting, contacting "FUNC" key automatically runs Span 

Calibration. When it suceeds, "SPAN SUCCESS" is displayed on LCD display for 2 sec then it changes 

to "CAL DATA" Mode.

- When span calibration is not successful, "SPAN FAIL" is displayed for 2 sec and it changes to 

"Cal. Data Mode".

SPAN >>

[  FAIL  ]

CAL.DATA

[      0]

- It is a mode to display measurement after calibration and checks whether calibration is successful after 

performing auto calibration.

- Contact "RESET" key to return to "Calibration mode".

7. System Mode

7.3. ALARM MODE

PASSWORD

[**]

- Contacting “FUNC” key for 2 sec or longer in Measuring Mode enters Password mode.
- After setting Password using "↑" or  "↓" key, contact "FUNC" key.

ALARM

MODE

- Contact "↑" or "↓" key to select "Alarm Mode".
- Contact "FUNC" key when "ALARM MODE" is displayed to enter Alarm setting mode.
- Contact "RESET" key to return to Measuring Mode.

ALM USED

[ ON]

- It is a setting where Alarm Mode Setting is turned on or off.

- Contact "↑" or "↓" key to change ON/OFF status. When it is ON, Alarm function and Alarm Mode 

setting can be performed.

- GTL-100 explosion-proof warning light can only be used when it is ON.

- Contacting "FUNC" key enters the next mode.

AL LATCH

[ ON]

- It is a mode that sets Reset method after running of Alarm1. Contact "↑" or "↓" key to change "ON" and 

"OFF".

- "OFF" sets to automatically reset the alarm. "ON" set to reset the alarm only when Reset key is pressed.
- Contact "FUNC" key when a desired mode is displayed to set then enter the next alarm setting mode.

A1 LEVEL

[    20]

- It is a mode that sets Alarm1 level. Contact "↑" or "↓" key to increase or decrease Alarm1 level, 

respectively.

- When a desired alarm 1 threshold is displayed, contact "FUNC" key to set it and enter the next item.

A1 TYPE

[INC]

- It is a mode that sets operational direction of Alarm1. Contact "↑" or "↓" key to display "INC" or "DEC".
- "INC" mode operates when the value is equal or larger than set alarm threshold. "DEC" mode operates 

when the value is equal or less than set alarm threshold.

- Contact "FUNC" key when a desired mode is displayed to set then enter the next mode.

A1 TYPE

[DEC]

A1 DBAND

[   0]%F

- It is a mode that sets Dead band value for Alarm1 operation. Use "↑" or "↓" key to set a value.

- It is a function where Alarm1 operates at values above the sum of Alarm1 level and dead band values 

and releases at values below Dead band value subtracted from Alarm1 level.

- When a desired alarm 1 dead band is displayed, press "FUNC" KEY to set it and enter the next item.

A1 TIME

    0 SEC

- It is a function to prevent instantaneous malfunction of detector due to external shock and noise other 

than from normal operation and time can be set in a range between 0~60 sec.

- For Alarm1 dwell time setting, press "↑" or "↓" key to increase or decrease in unit of 1 sec, respectively.
- When a desired Alarm 1 dwell time is displayed, press "FUNC" KEY to set it and enter the next item.
- Ex.) Alarm threshold value: 20% LEL / Delay time: When it is at 5 sec, Alarm triggers when the 

measured value is above the set value based on 20%LEL for 5 sec or longer. When it goes down below 

the set value within 5 sec, alarm is not triggered.

A2 LEVEL

[    20]

- It is a mode that sets Alarm2 level. Contact "↑" or "↓" key to increase or decrease Alarm2 level, respectively.
- When a desired alarm 2 threshold is displayed, press "FUNC" KEY to set it and enter the next item.

Summary of Contents for GTD-1000T Series

Page 1: ...dvance d Industry 1 ro Bugok dong Gunpo si Gyeonggi do Tel 82 31 490 0800 Fax 82 31 490 0801 Yeongnam business office Plant 55 Gonghangap gil 85beon gil Gangseogu Busan Metropolitan City Tel 51 973 8518 Fax 51 973 8519 E mail info gastron com www gastron com Read in detail for correct use ...

Page 2: ...e a responsibility and give you satisfaction In the present instruction manual operation method for Gas detector as well as simple methods for maintenance and repair etc are recorded If you read it in detail and keep it well for reference when you have questions then it will give you much help The present product and the product manual can be changed without advance notice for performance improvem...

Page 3: ...utions during Installation 28 12 Revision History 30 1 Overview 6 2 Configuration 6 3 Specification 7 3 1 Basic Specifications 7 3 2 Mechanical Specifications 7 3 3 Electrical Specifications Standard Type 8 3 4 Environmental Specifications 8 4 Name and Description of Each Part 9 4 1 Components 9 5 Installation 11 5 1 Detachment of Housing Cover 11 5 2 Main PCB Configuration 12 5 3 Terminal Configu...

Page 4: ...n perform calibration using detachable display LCD or OLED therefore maintenance is convenient 2 Configuration 1 Overview GTD 1000Tx M Instruction Manual Figure 1 GTD 1000Tx M Overview Gas Sensor Module GSA860EX Main Board Transmitter 24V DC 4 20mA Analog Signal Output Transmitter enclosure Safety Power Supply Unit 24VDC Safety Controller PLC or DCS GTD 1000TX M Gas Detector Hazardous Area Filed S...

Page 5: ...r velocity 6m s 3 Specification GTD 1000Tx M Instruction Manual 4 1 Components Figure 2 GTD 1000Tx M Components NO NAME DESCRIPTIONS 1 Housing Body Protects PCB Board built in Sensor and Housing from external environmental change and shock 2 Housing Cover It is assembled with detector housing body and protects PCB Board built in Sensor and Housing from external environmental change and shock 3 Sen...

Page 6: ...x M Instruction Manual Figure 3 Slotted Set Screw 5 Installation It is prohibited for an individual other than an approved user or a technician responsible for installation and repair from the head office to install a gas leak sensor on site or open the cover of the installed gas leak sensor and manipulate it This may cause serious loss of life and property from fire explosion and etc In addition ...

Page 7: ... CN3 Warning light GTL 100 Interface Connector 5 3 Terminal Configuration Warning Turn off power before connecting power terminal Loosen 5 terminal fixing screws located at top part of detached terminal block CN6 VIS mA ETH Connector by turning counter clockwise using a Θ driver Connect DC 18 24 V power to and then connect signal cable to mA Tighten 5 terminal fixing screws clockwise cw so that te...

Page 8: ... terminals for 4 20 mA sink output at PLC side to VISO terminal and mA terminal respectively Then turn off the J1 jumper The maximum length between GTD 1000Tx M and power supply is decided by wire specification Max Installation Length VMAXDROP IMAX WIRER m 2 VMAXDROP Maximum Power Loop Voltage Drop Power Supply voltage min operating voltage IMAX Max Current of GTD 1000Tx M WIRER m The resistance o...

Page 9: ...d row the current measurement is displayed In the current screen touching the magnet bar on Func key for 2 sec or longer changes to setting mode During Modbus network is displayed on the left of the first row ENG When the mode is on is displayed on the left of the second row COMB 0 0 LEL 30 When alarm1 or 2 occurs it runs as below It runs only when alarm used setting in alarm mode is on The first ...

Page 10: ...ORD 00 99 00 CALIBRA MODE Calibration Mode CALIBRA ZERO Calibration Zero ZERO CAL NO YES NO ZERO GAS 0 ZERO SUCCESS FAIL CAL DATA 0 FAIL CALIBRA SPAN Calibration Span SPAN CAL NO YES NO SPAN GAS 0 SPAN SET 50 FAIL 50 SPAN SUCCESS FAIL CAL DATA 0 ALARM MODE ALM USED Alarm Used OFF ON OFF AL LATCH Alarm Latch OFF ON OFF A1 LEVEL Alarm 1 Level 1 10000 20 A1 TYPE Alarm 1 Type INC DEC INC A1 DBAND Alar...

Page 11: ...desired high scale is displayed contact FUNC key to set the high scale value and enter the next Program Mode PASSWORD 00 It is password setting mode This password is checked when entering Program Mode Set using or key then contact FUNC key to set the password and enter the next Program Mode 7 System Mode 7 System Mode 7 2 Calibration Mode 7 2 1 Zero Calibration PASSWORD Contacting FUNC key for 2 s...

Page 12: ...MODE is displayed to enter Alarm setting mode Contact RESET key to return to Measuring Mode ALM USED ON It is a setting where Alarm Mode Setting is turned on or off Contact or key to change ON OFF status When it is ON Alarm function and Alarm Mode setting can be performed GTL 100 explosion proof warning light can only be used when it is ON Contacting FUNC key enters the next mode AL LATCH ON It is...

Page 13: ...g or key contact FUNC key DEVICE MODE Contact or key to select CALIBRA MODE When DEVICE MODE is displayed contact FUNC key to enter Device Mode Contact RESET key to return to Measuring Mode 485 B D CHECK It is a setting where connection status of RS 485 Board is confirmed When it is connected it displays CHECK If not it displays EMPTY ADDRESS OFF It is a setting where RS 485 Modbus Address is set ...

Page 14: ...nt is not considered Alarm 2 Setting 30004 BIT15 0 Alarm 2 Setting Integer Decimal point is not considered Gas detector status value 10000 BIT0 Alarm 1 Active Status BIT1 Alarm 2 Active Status BIT2 Fault Active Status BIT3 Maintenance Mode Status BIT4 Test Mode Status BIT5 Calibration Mode Status BIT6 Reserved BIT7 Toggle Bit Bit reversal in 2 sec interval External Test 3 BIT0 7 Gas Detector Test ...

Page 15: ...s during Installation Avoid areas with electrical barriers such as rain water etc It is recommended to be installed in areas that are easier to work in since regular maintenance is needed Avoid areas with vibration or shock since they can affect output values Sensor part must be installed towards the direction of gravity This equipment has explosion proof construction for internal pressure and bel...

Page 16: ...www gastron com 30_31 GTD 1000Tx M Instruction Manual 12 Revision History VERSION CONTENTS DATE 1 0 Manual Initial Revision 2017 11 17 1 1 Changed Address 2018 01 04 ...

Reviews: