GPD-100 Instruction Manual
PAGE 17 of 27
Rev 1.1(2019.05.23)
device goes to next item.
This is a password mode in which you obtain an authority to change the
program of detector. Password increases or decreases whenever you press
“
” key or “
” key.
When desired password is displayed, press “RUN/MODE” key. Then a
password is set to be the number displayed in the screen. After that, the
device goes to next item.
This is a suppression mode(“SUPr”) in which you can set suppression
percentage of gas concentration that is supposed to be presented as ‘0’.
The percentage number increases or decreases whenever “
” key or “
”
key is pressed. (Possible to set OFF or 1~50)
When desired number is displayed, press “RUN/MODE” key. Then a
suppression percentage is set to be the number displayed. After that, the
device goes to next item.
This is a Pyrolyzer(CEC) mode in which you can set whether Pyrolyzer is
applied or not. When you press “
” key or “
” key, it will switch
between ON and OFF alternatively.
When desired option is displayed, press “RUN/MODE” key. Then Pyrolyzer
mode is specified. After that, the device goes to next item.
This is a “Py-V” mode in which you can set the voltage level used by
Pyrolyzer.
When you press ”RUN/MODE” key, you will enter voltage setting function.
Pyrolyzer voltage is a mode in which you can set voltage value used by
pyrolyzer of detector. When you press “
” key or “
” key, the voltage
level( 1.50V ) will increase or decrease.
Voltage can be set between 0.5V and 2.3V. When desired value is
displayed, press “RUN/MODE” key. Then the voltage is set to the value
displayed. After that, the device goes to next item.
This is a mode in which program version is viewed.
If you press “RUN/MODE” key, you will enter a menu mode.