Warnings and Cautions 

A. Signal 

Words 

The following signal words, as defined by ANSI Z535.4-1998, are used in the ToxiPro 
Operator’s Guide. 

 indicates an imminently hazardous situation which, if not avoided, 

will result in death or serious injury. 

 indicates a potentially hazardous situation which, if not avoided, 

could result in death or serious injury. 

 indicates a potentially hazardous situation, which if not avoided, may 

result in moderate or minor injury. 

CAUTION

 used without the safety alert symbol indicates a potentially hazardous 

situation which, if not avoided, may result in property damage. 

B. Warnings 
1. 

 ToxiPro personal, portable gas detectors have been designed for 

the detection of either oxygen deficiencies or specific toxic gas accumulations.  An 
alarm condition indicates the presence of a potentially life-threatening hazard and 
should be taken very seriously. 

2. 

 In the event of an alarm condition it is important to follow 

established procedures.  The safest course of action is to immediately leave the 
affected area, and to return only after further testing determines that the area is 
once again safe for entry.  Failure to immediately leave the area may result in 
serious injury or death. 

3. 

 Use only Duracell Ultra #CR2, Energizer #1CR2 (EL1CR2), Sanyo 

#CR2 or Panasonic #CR2 in the ToxiPro.

 

4. 

 The accuracy of ToxiPro instruments equipped with toxic gas 

sensors should be checked periodically with known concentration calibration gas.  
Failure to check accuracy can lead to inaccurate and potentially dangerous 
readings.  The ToxiPro O

2

 should be periodically calibrated in fresh air. 

5. 

 A sensor that cannot be calibrated or is found to be out of 

tolerance must be replaced immediately.  An instrument equipped with a toxic gas 
sensor that fails calibration may not be used until testing with known concentration 
test gas determines that accuracy has been restored, and the instrument is once 
again fit for use.  Instruments equipped with an oxygen sensor that fail calibration 
may not be used until testing with fresh air determines that accuracy has been 
restored and the instrument is once again fit for use. 

6. 

 Do not reset the calibration gas concentration setpoints in the 

ToxiPro unless the concentrations of your calibration gas differ from the 
concentrations of the calibration gas that is normally supplied by Sperian 
Instrumentation for use in calibrating the ToxiPro. 

7. 

  Use of non-standard calibration gas and/or calibration kit 

components when calibrating the ToxiPro can lead to dangerously inaccurate 
readings and may void the standard Sperian Instrumentation warranty. 

 

Sperian Instrumentation offers calibration kits and long-lasting cylinders of test gas 
specifically developed for easy calibration.  Customers are strongly urged to use 
only Sperian Instrumentation calibration materials when calibrating the ToxiPro. 

8. 

  Substitution of components may impair intrinsic safety. 

9. 

  For safety reasons the ToxiPro must be operated by qualified 

personnel only.  Read, understand and follow the directions set forth in this 
reference manual before operating the ToxiPro. 

10. 

  The ToxiPro has been tested for intrinsic safety in Explosive 

Gas/AIR (max. 21.0% O

2

). 

Summary of Contents for ToxiPro

Page 1: ...ToxiPro S ing le G as Detec tor R eferenc e M anual GasTech Australia Pty Ltd 24 Baretta Rd Wangara Western Australia 6065 Tel 1800 999 902 Fax 1800 999 903 http www gastech com au ...

Page 2: ...E INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ToxiPro Reference Manual Sperian Instrumentation Part Number 13 264 Version 2 01 Copyright 2008 by Sperian Protection Instrumentation LLC Middletown Connecticut 06457 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Sp...

Page 3: ...kout 8 2 2 Backlight 8 2 3 Functions 8 2 4 Calibration and bump test due notices 8 2 5 Low battery alarms 9 2 6 Turning the ToxiPro off 9 2 7 Always On Mode 9 2 7 1 Turning Off in Always On Mode 9 2 8 Sampling 9 2 8 1 Sample draw kit usage 9 3 Calibration 10 3 1 Verification of accuracy 10 3 1 1 Verifying accuracy and response ToxiPro O2 10 3 1 2 Verifying accuracy ToxiPro with toxic gas sensor 10...

Page 4: ...erian Instrumentation Warranty Gas Detection Products 18 Operating Temperature Range The safe operating temperature range of the gas detector is printed on the back label of the instrument Use of Sperian Gas Detectors outside of the instrument s specified operating temperature range may result in inaccurate and potentially dangerous readings Certifications The ToxiPro carries the following certifi...

Page 5: ...accurate and potentially dangerous readings The ToxiPro O2 should be periodically calibrated in fresh air 5 A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrume...

Page 6: ... B 1 2 2 ToxiPro O2 sensor ranges The oxygen sensor used in the ToxiPro O2 has a range of 0 30 by volume 1 2 3 ToxiPro O2 warm up phase The ToxiPro O2 requires a one time 15 minute warm up phase prior to initial activation See section 2 1 for further details 1 3 Alarm and warning logic 1 3 1 Gas alarms ToxiPro gas alarms are user adjustable and may be set anywhere within the range of the specific ...

Page 7: ...n A built in manually activated backlight allows the display to be read even in low light conditions Visual alarm LED IrDA port A bright red LED Light Emitting Diode alarm light provides a visual indication of the alarm state The LED also functions as the IrDA port MODE button The large push button on the front of the instrument is called the MODE button The MODE button is used to turn the ToxiPro...

Page 8: ...hat the 6 digit serial number can not fit on a single screen so it is shown on two screens In this case the instrument serial number is 481098 Datalogging versions will then show the dL screen Once the sensor is recognized the instrument will display the sensor type The warning alarm level will then be displayed followed by the danger alarm level During the display of the warning alarm level the L...

Page 9: ...the MAX screen press the MODE button to view the time screen A dot below the colon between the hours and the minutes digits indicates afternoon evening hours pm 10 45 am 10 45 pm If the STEL Short Term Exposure Limit alarm is enabled press the MODE button once to view the STEL reading The STEL value displayed represents the average value of the instrument readings for the target gas for the most r...

Page 10: ...re Note The ToxiPro must be turned off to replace the battery See section 2 7 1 2 7 1 Turning Off in Always On Mode BioTrak software must be used to turn the ToxiPro off when it is in Always on Mode BioTrak is available at http www biodownloads com 1 Open BioTrak and select ToxiPro 2 Select Configure in the software 3 Hold the MODE button down for about 10 seconds until IrDA is shown and establish...

Page 11: ...ds then slowly exhale directly onto the face of the sensor in the same way you would to fog up a piece of glass If the descending oxygen alarm is set to 19 5 the instrument should go into alarm after a few seconds 3 1 2 Verifying accuracy ToxiPro with toxic gas sensor Verification of accuracy is a two step procedure for ToxiPro instruments equipped with a toxic gas sensor Step one is to take the T...

Page 12: ... between the following two screens 3 For instruments equipped with a toxic gas sensor the fresh air zero calibration is complete when the instrument begins a second 5 second countdown for the span calibration If span calibration is not required allow the countdown to reach 0 without pressing the MODE button For further instructions concerning the span calibration of toxic sensor equipped models pr...

Page 13: ...ibration is fully automatic from this point on Once the instrument successfully completes the span calibration it will emit three short beeps and display the maximum span calibration adjustment value for two seconds Note The maximum span calibration adjustment value shown is an indication of the relative health of the sensor As a sensor loses sensitivity the maximum adjustment level will approach ...

Page 14: ...an calibration failure in which calibration gas is not detected the no and GAS screens will be alternately displayed as shown below with the calibration bottle icon highlighted When the instrument returns to the current gas readings screen the warning symbol and the calibration bottle icon will be shown which signifies that the last attempt to span calibrate the instrument has failed Note If the C...

Page 15: ...utomatically restart with the hours digits blinking The next five steps describe programming the time and date 7 Press and release the MODE button to advance the setting by one Hold the MODE button down to scroll rapidly Afternoon and evening hours are differentiated by a dot below the colon on the display 10 a m 10 p m Once the appropriate hour setting is shown wait 5 seconds and the minutes sett...

Page 16: ... instrument functionality 4 4 Storage ToxiPro detectors may be stored for long periods in a fresh air environment at temperatures between 10 C 50 F and 30 C 86 F 5 PC Instrument Communications 5 1 Event logging Each ToxiPro includes a built in event logger that stores instrument readings during alarm conditions calibration values and other data BioTrak software and an IrDA port are necessary to do...

Page 17: ... 50 150 0 2 n d n d 0 Nitrogen Dioxide NO2 5 8 1 0 100 90 270 0 n d n d 0 n d n d Chlorine Cl2 specific 0 3 0 n d 12 100 20 0 0 0 0 0 0 Chlorine Dioxide ClO2 specific 0 25 0 n d n d 60 100 0 0 0 n d 0 0 Ammonia NH3 54 47 21 0 5 0 n d 0 0 n d 0 0 0 100 0 0 Phosphine PH3 0 5 25 20 n d 0 1 n d n d n d 1 0 5 Hydrogen Cyanide HCN 0 5 350 160 5 100 20 60 0 1 100 65 5 50 n d n d no data undetermined posi...

Page 18: ...mmends the following procedure to establish a safe and prudent accuracy check schedule for your Sperian instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that the sensors are exposed...

Page 19: ... is in service ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Toxi3Ltd 3 years after activation or 3 years after the Must Be Activated By date whichever comes first Mighty Tox 2 Prorated credit is given towards repair or purchase of a new unit of the same type 0 6 months of use 100 credit 6 12 mon...

Page 20: ...xiPro c UL Version ATEX CQST Version IECEx Version Type O2 H2S SO2 CO CO H CO NH3 Cl2 ClO2 NO2 HCN PH3 SO2 Hi Software Version _____ Vibrating Motor Datalogging Belt Clip Serial Number ____________ MFG Date ...

Reviews: