background image

 

10

remote location to the sensor 
compartment.  Allow one squeeze of 
the bulb for every one foot of sampling 
hose for the sample to reach the 
sensors.  Continue to squeeze the bulb 
for an additional 45 seconds or until 
readings stabilize. 

6.  Note the gas measurement readings.

 

CAUTION:  Hand aspirated remote 
sampling only provides continuous gas 
readings for the area in which the probe is 
located when the bulb is being 
continuously squeezed.   
Note:  Each time a reading is desired, it is 
necessary to squeeze the bulb a sufficient 
number of times to bring a fresh sample to 
the sensor compartment and to continue 
to squeeze the bulb until readings 
stabilize.

 

3. Calibration 

The ToxiPro features fully automated fresh 
air/zero and span calibration functions.  The 
MODE button is used to initiate the automatic 
calibration sequence.  Calibration 
adjustments are made automatically by the 
instrument. 

3.1 

Verification of accuracy 

3.1.1  Verifying accuracy and response: 

ToxiPro O

2

 

To verify the accuracy of the ToxiPro O

2

, take 

the ToxiPro O

2

 to an area where the 

atmosphere is known to be fresh and check 
the readings.  If the readings differ from those 
expected in fresh air (oxygen monitors should 
read 20.9% in fresh air), then a fresh air/zero 
calibration adjustment must be made as 
discussed below in section 3.3.  If fresh air is 
not available, see section 3.7 below for 
instructions for calibrating the ToxiPro O

2

 in 

contaminated air. 
Sperian Instrumentation also recommends 
that the response of the oxygen sensor be 
regularly verified by any of these methods: 

  Expose the O

2

 sensor to a known 

concentration of gas containing less than 
19.0% oxygen.  If the descending oxygen 
alarm is set to 19.5% the instrument 
should go into alarm a few seconds after 
the gas reaches the sensor face. 

  Process the ToxiPro O

in an IQ Express 

Dock that has a cylinder of calibration gas 
containing 18.0% oxygen (or less) 
connected to the gas port and proceed 
with the oxygen bump test.  

  Breath test:  Hold your breath for 10 

seconds, then 

slowly

 exhale directly onto 

the face of the sensor (in the same way 

you would to fog up a piece of glass).  If 
the descending oxygen alarm is set to 
19.5%, the instrument should go into 
alarm after a few seconds. 

3.1.2  Verifying accuracy: ToxiPro with 

toxic gas sensor 

Verification of accuracy is a two-step 
procedure for ToxiPro instruments equipped 
with a toxic gas sensor: 
Step one is to take the ToxiPro to an area 
where the atmosphere is known to be fresh 
and check the readings.  If the readings differ 
from those expected in fresh air (instruments 
equipped with a toxic sensor should read 0 
PPM in fresh air), then a fresh air calibration 
adjustment must be made as discussed 
below in section 3.3. 
Step two is to test sensor response by 
exposing the sensor to a test gas of known 
concentration.  This is known as a functional 
(bump) test.  Readings are considered to be 
accurate when the display is between 90% 
and 120% of the expected values as given on 
the calibration gas cylinder.  If readings are 
accurate, there is no need to adjust your gas 
detector.  See section 3.4 for further details 
concerning the functional/bump test. 
If the readings are inaccurate, the instrument 
must be span calibrated before further use as 
discussed in section 3.5. 

  

The accuracy of 

ToxiPro instruments equipped with toxic 
gas sensors should be checked 
periodically with known concentration 
calibration gas.  Failure to check accuracy 
can lead to inaccurate and potentially 
dangerous readings. 
See Appendix C for a discussion of 
calibration frequency recommendations.

 

3.2 

Effect of contaminants on ToxiPro 

sensors 

The atmosphere in which the ToxiPro is used 
can have lasting effects on the sensors.  
Sensors may suffer losses in sensitivity 
leading to degraded performance if exposed 
to certain substances. 
The ToxiPro O

2

 uses a galvanic oxygen 

sensor, while toxic sensor versions of the 
ToxiPro use an electrochemical toxic gas 
sensor.  Different types of sensors use 
different detection principles, so the 
conditions that affect the accuracy of the 
sensors vary from one type of sensor to the 
next. 

3.2.1  Effects of contaminants on O

2

 

sensors 

Oxygen sensors may be affected by 
prolonged exposure to "acid" gases such as 

Summary of Contents for ToxiPro

Page 1: ...ToxiPro S ing le G as Detec tor R eferenc e M anual GasTech Australia Pty Ltd 24 Baretta Rd Wangara Western Australia 6065 Tel 1800 999 902 Fax 1800 999 903 http www gastech com au ...

Page 2: ...E INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ToxiPro Reference Manual Sperian Instrumentation Part Number 13 264 Version 2 01 Copyright 2008 by Sperian Protection Instrumentation LLC Middletown Connecticut 06457 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Sp...

Page 3: ...kout 8 2 2 Backlight 8 2 3 Functions 8 2 4 Calibration and bump test due notices 8 2 5 Low battery alarms 9 2 6 Turning the ToxiPro off 9 2 7 Always On Mode 9 2 7 1 Turning Off in Always On Mode 9 2 8 Sampling 9 2 8 1 Sample draw kit usage 9 3 Calibration 10 3 1 Verification of accuracy 10 3 1 1 Verifying accuracy and response ToxiPro O2 10 3 1 2 Verifying accuracy ToxiPro with toxic gas sensor 10...

Page 4: ...erian Instrumentation Warranty Gas Detection Products 18 Operating Temperature Range The safe operating temperature range of the gas detector is printed on the back label of the instrument Use of Sperian Gas Detectors outside of the instrument s specified operating temperature range may result in inaccurate and potentially dangerous readings Certifications The ToxiPro carries the following certifi...

Page 5: ...accurate and potentially dangerous readings The ToxiPro O2 should be periodically calibrated in fresh air 5 A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrume...

Page 6: ... B 1 2 2 ToxiPro O2 sensor ranges The oxygen sensor used in the ToxiPro O2 has a range of 0 30 by volume 1 2 3 ToxiPro O2 warm up phase The ToxiPro O2 requires a one time 15 minute warm up phase prior to initial activation See section 2 1 for further details 1 3 Alarm and warning logic 1 3 1 Gas alarms ToxiPro gas alarms are user adjustable and may be set anywhere within the range of the specific ...

Page 7: ...n A built in manually activated backlight allows the display to be read even in low light conditions Visual alarm LED IrDA port A bright red LED Light Emitting Diode alarm light provides a visual indication of the alarm state The LED also functions as the IrDA port MODE button The large push button on the front of the instrument is called the MODE button The MODE button is used to turn the ToxiPro...

Page 8: ...hat the 6 digit serial number can not fit on a single screen so it is shown on two screens In this case the instrument serial number is 481098 Datalogging versions will then show the dL screen Once the sensor is recognized the instrument will display the sensor type The warning alarm level will then be displayed followed by the danger alarm level During the display of the warning alarm level the L...

Page 9: ...the MAX screen press the MODE button to view the time screen A dot below the colon between the hours and the minutes digits indicates afternoon evening hours pm 10 45 am 10 45 pm If the STEL Short Term Exposure Limit alarm is enabled press the MODE button once to view the STEL reading The STEL value displayed represents the average value of the instrument readings for the target gas for the most r...

Page 10: ...re Note The ToxiPro must be turned off to replace the battery See section 2 7 1 2 7 1 Turning Off in Always On Mode BioTrak software must be used to turn the ToxiPro off when it is in Always on Mode BioTrak is available at http www biodownloads com 1 Open BioTrak and select ToxiPro 2 Select Configure in the software 3 Hold the MODE button down for about 10 seconds until IrDA is shown and establish...

Page 11: ...ds then slowly exhale directly onto the face of the sensor in the same way you would to fog up a piece of glass If the descending oxygen alarm is set to 19 5 the instrument should go into alarm after a few seconds 3 1 2 Verifying accuracy ToxiPro with toxic gas sensor Verification of accuracy is a two step procedure for ToxiPro instruments equipped with a toxic gas sensor Step one is to take the T...

Page 12: ... between the following two screens 3 For instruments equipped with a toxic gas sensor the fresh air zero calibration is complete when the instrument begins a second 5 second countdown for the span calibration If span calibration is not required allow the countdown to reach 0 without pressing the MODE button For further instructions concerning the span calibration of toxic sensor equipped models pr...

Page 13: ...ibration is fully automatic from this point on Once the instrument successfully completes the span calibration it will emit three short beeps and display the maximum span calibration adjustment value for two seconds Note The maximum span calibration adjustment value shown is an indication of the relative health of the sensor As a sensor loses sensitivity the maximum adjustment level will approach ...

Page 14: ...an calibration failure in which calibration gas is not detected the no and GAS screens will be alternately displayed as shown below with the calibration bottle icon highlighted When the instrument returns to the current gas readings screen the warning symbol and the calibration bottle icon will be shown which signifies that the last attempt to span calibrate the instrument has failed Note If the C...

Page 15: ...utomatically restart with the hours digits blinking The next five steps describe programming the time and date 7 Press and release the MODE button to advance the setting by one Hold the MODE button down to scroll rapidly Afternoon and evening hours are differentiated by a dot below the colon on the display 10 a m 10 p m Once the appropriate hour setting is shown wait 5 seconds and the minutes sett...

Page 16: ... instrument functionality 4 4 Storage ToxiPro detectors may be stored for long periods in a fresh air environment at temperatures between 10 C 50 F and 30 C 86 F 5 PC Instrument Communications 5 1 Event logging Each ToxiPro includes a built in event logger that stores instrument readings during alarm conditions calibration values and other data BioTrak software and an IrDA port are necessary to do...

Page 17: ... 50 150 0 2 n d n d 0 Nitrogen Dioxide NO2 5 8 1 0 100 90 270 0 n d n d 0 n d n d Chlorine Cl2 specific 0 3 0 n d 12 100 20 0 0 0 0 0 0 Chlorine Dioxide ClO2 specific 0 25 0 n d n d 60 100 0 0 0 n d 0 0 Ammonia NH3 54 47 21 0 5 0 n d 0 0 n d 0 0 0 100 0 0 Phosphine PH3 0 5 25 20 n d 0 1 n d n d n d 1 0 5 Hydrogen Cyanide HCN 0 5 350 160 5 100 20 60 0 1 100 65 5 50 n d n d no data undetermined posi...

Page 18: ...mmends the following procedure to establish a safe and prudent accuracy check schedule for your Sperian instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that the sensors are exposed...

Page 19: ... is in service ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Toxi3Ltd 3 years after activation or 3 years after the Must Be Activated By date whichever comes first Mighty Tox 2 Prorated credit is given towards repair or purchase of a new unit of the same type 0 6 months of use 100 credit 6 12 mon...

Page 20: ...xiPro c UL Version ATEX CQST Version IECEx Version Type O2 H2S SO2 CO CO H CO NH3 Cl2 ClO2 NO2 HCN PH3 SO2 Hi Software Version _____ Vibrating Motor Datalogging Belt Clip Serial Number ____________ MFG Date ...

Reviews: