background image

 

17

Appendix C:  Calibration 
Frequency 

Recommendation 

One of the most common 
questions that we are asked at 
Sperian Instrumentation is: 

“How 

often should I calibrate my gas 
detector?”

 

Sensor Reliability and Accuracy 

Today’s sensors are designed to 
provide years of reliable service.  In 
fact, many sensors are designed 
so that with normal use they will 
only lose 5% of their sensitivity per 
year or 10% over a two-year 
period.  Given this, it should be 
possible to use a sensor for up to 
two full years without significant 

loss of sensitivity. 

Verification of Accuracy 

With so many reasons why a 
sensor can lose sensitivity and 
given the fact that dependable 
sensors can be key to survival in a 
hazardous environment, frequent 
verification of sensor performance 

is paramount. 
There is only one sure way to verify 
that a sensor can respond to the 
gas for which it is designed. That is 
to expose it to a known 
concentration of target gas and 
compare the reading with the 
concentration of the gas.  This is 
referred to as a “bump” test. This 
test is very simple and takes only a 
few seconds to accomplish. 

The 

safest course of action is to do a 
“bump” test prior to each day’s 
use

. It is not necessary to make a 

calibration adjustment if the 
readings fall between 90%* and 
120% of the expected value. As an 
example, if a CO sensor is checked 
using a gas concentration of 50 
PPM it is not necessary to perform 
a calibration unless the readings 
are either below 45 PPM or above 

60 PPM. 

*The Canadian Standards 
Association (CSA) requires 
combustible gas sensors to 
undergo calibration when the 
displayed value during a bump 
test fails to fall between 100% 
and 120% of the expected value 
for the gas. 

Lengthening the Intervals 

between Verification of Accuracy 

We are often asked whether there 
are any circumstances in which the 
period between accuracy checks 
may be lengthened.  

Sperian Instrumentation is not the 
only manufacturer to be asked this 
question! One of the professional 
organizations to which Sperian 
Instrumentation belongs is the 
Industrial Safety Equipment 
Association (ISEA). The 
“Instrument Products” group of this 
organization has been very active 
in developing a protocol to clarify 
the minimum conditions under 
which the interval between 
accuracy checks may be 

lengthened. 
A number of leading gas detection 
equipment manufacturers have 
participated in the development of 
the ISEA guidelines concerning 
calibration frequency. Sperian 
Instrumentation’s procedures 
closely follow these guidelines.  
If your operating procedures do not 
permit daily checking of the 
sensors, Sperian Instrumentation 
recommends the following 
procedure to establish a safe and 
prudent accuracy check schedule 

for your Sperian instruments: 
1.  During a period of initial use of 

at least 10 days in the 
intended atmosphere, check 
the sensor response daily to 
be sure there is nothing in the 
atmosphere that is poisoning 
the sensor(s). The period of 
initial use must be of sufficient 
duration to ensure that the 
sensors are exposed to all 
conditions that might have an 

adverse effect on the sensors.  

2.  If these tests demonstrate that 

it is not necessary to make 
adjustments, the time between 
checks may be lengthened. 
The interval between accuracy 
checking should not exceed 30 

days.  

3. 

When the interval has been 
extended the toxic and 
combustible gas sensors 
should be replaced 
immediately upon warranty 
expiration. This will minimize 
the risk of failure during the 
interval between sensor 

checks. 

 

4.  The history of the instrument 

response between verifications 
should be kept. Any 
conditions, incidents, 
experiences, or exposure to 
contaminants that might have 
an adverse effect on the 
calibration state of the sensors 
should trigger immediate re-

verification of accuracy before 

further use.  

5.  Any changes in the 

environment in which the 
instrument is being used, or 
changes in the work that is 
being performed, should 
trigger a resumption of daily 

checking.  

6.  If there is any doubt at any 

time as to the accuracy of the 
sensors, verify the accuracy of 
the sensors by exposing them 
to known concentration test 
gas before further use.  

Gas detectors used for the 
detection of oxygen deficiencies, 
flammable gases and vapors, or 
toxic contaminants must be 
maintained and operated properly 
to do the job they were designed to 
do. Always follow the guidelines 
provided by the manufacturer for 
any gas detection equipment you 

use! 
If there is any doubt regarding your 
gas detector's accuracy, do an 
accuracy check! All it takes is a few 
moments to verify whether or not 

your instruments are safe to use. 

 

One Button Auto Calibration 

While it is only necessary to do a 
“bump” test to ensure that the 
sensors are working properly, all 
current Sperian gas detectors offer 
a one-button auto calibration 
feature. This feature allows you to 
calibrate a Sperian gas detector in 
about the same time as it takes to 
complete a “bump” test. The use of 
automatic bump test and 
calibration stations can further 
simplify the tasks, while 

automatically maintaining records. 

Don't take a chance  

with your life.   

Verify accuracy frequently! 

Please read also Sperian 
Instrumentation’s application note: 

AN20010808 “Use of ‘equivalent’ 
calibration gas mixtures”

.  This 

application note provides 
procedures to ensure safe 
calibration of LEL sensors that are 
subject to silicone poisoning.

 

 

Sperian Instrumentation’s  website 
is located at   
 

http://www.biosystems.com 

Summary of Contents for ToxiPro

Page 1: ...ToxiPro S ing le G as Detec tor R eferenc e M anual GasTech Australia Pty Ltd 24 Baretta Rd Wangara Western Australia 6065 Tel 1800 999 902 Fax 1800 999 903 http www gastech com au ...

Page 2: ...E INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ToxiPro Reference Manual Sperian Instrumentation Part Number 13 264 Version 2 01 Copyright 2008 by Sperian Protection Instrumentation LLC Middletown Connecticut 06457 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Sp...

Page 3: ...kout 8 2 2 Backlight 8 2 3 Functions 8 2 4 Calibration and bump test due notices 8 2 5 Low battery alarms 9 2 6 Turning the ToxiPro off 9 2 7 Always On Mode 9 2 7 1 Turning Off in Always On Mode 9 2 8 Sampling 9 2 8 1 Sample draw kit usage 9 3 Calibration 10 3 1 Verification of accuracy 10 3 1 1 Verifying accuracy and response ToxiPro O2 10 3 1 2 Verifying accuracy ToxiPro with toxic gas sensor 10...

Page 4: ...erian Instrumentation Warranty Gas Detection Products 18 Operating Temperature Range The safe operating temperature range of the gas detector is printed on the back label of the instrument Use of Sperian Gas Detectors outside of the instrument s specified operating temperature range may result in inaccurate and potentially dangerous readings Certifications The ToxiPro carries the following certifi...

Page 5: ...accurate and potentially dangerous readings The ToxiPro O2 should be periodically calibrated in fresh air 5 A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrume...

Page 6: ... B 1 2 2 ToxiPro O2 sensor ranges The oxygen sensor used in the ToxiPro O2 has a range of 0 30 by volume 1 2 3 ToxiPro O2 warm up phase The ToxiPro O2 requires a one time 15 minute warm up phase prior to initial activation See section 2 1 for further details 1 3 Alarm and warning logic 1 3 1 Gas alarms ToxiPro gas alarms are user adjustable and may be set anywhere within the range of the specific ...

Page 7: ...n A built in manually activated backlight allows the display to be read even in low light conditions Visual alarm LED IrDA port A bright red LED Light Emitting Diode alarm light provides a visual indication of the alarm state The LED also functions as the IrDA port MODE button The large push button on the front of the instrument is called the MODE button The MODE button is used to turn the ToxiPro...

Page 8: ...hat the 6 digit serial number can not fit on a single screen so it is shown on two screens In this case the instrument serial number is 481098 Datalogging versions will then show the dL screen Once the sensor is recognized the instrument will display the sensor type The warning alarm level will then be displayed followed by the danger alarm level During the display of the warning alarm level the L...

Page 9: ...the MAX screen press the MODE button to view the time screen A dot below the colon between the hours and the minutes digits indicates afternoon evening hours pm 10 45 am 10 45 pm If the STEL Short Term Exposure Limit alarm is enabled press the MODE button once to view the STEL reading The STEL value displayed represents the average value of the instrument readings for the target gas for the most r...

Page 10: ...re Note The ToxiPro must be turned off to replace the battery See section 2 7 1 2 7 1 Turning Off in Always On Mode BioTrak software must be used to turn the ToxiPro off when it is in Always on Mode BioTrak is available at http www biodownloads com 1 Open BioTrak and select ToxiPro 2 Select Configure in the software 3 Hold the MODE button down for about 10 seconds until IrDA is shown and establish...

Page 11: ...ds then slowly exhale directly onto the face of the sensor in the same way you would to fog up a piece of glass If the descending oxygen alarm is set to 19 5 the instrument should go into alarm after a few seconds 3 1 2 Verifying accuracy ToxiPro with toxic gas sensor Verification of accuracy is a two step procedure for ToxiPro instruments equipped with a toxic gas sensor Step one is to take the T...

Page 12: ... between the following two screens 3 For instruments equipped with a toxic gas sensor the fresh air zero calibration is complete when the instrument begins a second 5 second countdown for the span calibration If span calibration is not required allow the countdown to reach 0 without pressing the MODE button For further instructions concerning the span calibration of toxic sensor equipped models pr...

Page 13: ...ibration is fully automatic from this point on Once the instrument successfully completes the span calibration it will emit three short beeps and display the maximum span calibration adjustment value for two seconds Note The maximum span calibration adjustment value shown is an indication of the relative health of the sensor As a sensor loses sensitivity the maximum adjustment level will approach ...

Page 14: ...an calibration failure in which calibration gas is not detected the no and GAS screens will be alternately displayed as shown below with the calibration bottle icon highlighted When the instrument returns to the current gas readings screen the warning symbol and the calibration bottle icon will be shown which signifies that the last attempt to span calibrate the instrument has failed Note If the C...

Page 15: ...utomatically restart with the hours digits blinking The next five steps describe programming the time and date 7 Press and release the MODE button to advance the setting by one Hold the MODE button down to scroll rapidly Afternoon and evening hours are differentiated by a dot below the colon on the display 10 a m 10 p m Once the appropriate hour setting is shown wait 5 seconds and the minutes sett...

Page 16: ... instrument functionality 4 4 Storage ToxiPro detectors may be stored for long periods in a fresh air environment at temperatures between 10 C 50 F and 30 C 86 F 5 PC Instrument Communications 5 1 Event logging Each ToxiPro includes a built in event logger that stores instrument readings during alarm conditions calibration values and other data BioTrak software and an IrDA port are necessary to do...

Page 17: ... 50 150 0 2 n d n d 0 Nitrogen Dioxide NO2 5 8 1 0 100 90 270 0 n d n d 0 n d n d Chlorine Cl2 specific 0 3 0 n d 12 100 20 0 0 0 0 0 0 Chlorine Dioxide ClO2 specific 0 25 0 n d n d 60 100 0 0 0 n d 0 0 Ammonia NH3 54 47 21 0 5 0 n d 0 0 n d 0 0 0 100 0 0 Phosphine PH3 0 5 25 20 n d 0 1 n d n d n d 1 0 5 Hydrogen Cyanide HCN 0 5 350 160 5 100 20 60 0 1 100 65 5 50 n d n d no data undetermined posi...

Page 18: ...mmends the following procedure to establish a safe and prudent accuracy check schedule for your Sperian instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that the sensors are exposed...

Page 19: ... is in service ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Toxi3Ltd 3 years after activation or 3 years after the Must Be Activated By date whichever comes first Mighty Tox 2 Prorated credit is given towards repair or purchase of a new unit of the same type 0 6 months of use 100 credit 6 12 mon...

Page 20: ...xiPro c UL Version ATEX CQST Version IECEx Version Type O2 H2S SO2 CO CO H CO NH3 Cl2 ClO2 NO2 HCN PH3 SO2 Hi Software Version _____ Vibrating Motor Datalogging Belt Clip Serial Number ____________ MFG Date ...

Reviews: