19
Enter your User ID and password again
when the Security Login window is
shown. A window will be shown to
indicate that the dock configuration has
been updated.
11. Click OK. “PIN Lock Enabled” will
then be shown in the PC software in
the dock configuration window.
The dock will now verify the PIN in any
instrument that interfaces with it.
12. For every other dock in the system,
access the dock’s advanced settings,
enter the PIN, click on “Check PIN”
and click the “Set Config” button.
With PIN lock enabled, only instruments
with a PIN matching the dock’s PIN will
be processed.
4. Dock
Use
Once the IQ6 is connected to the gas
cylinder(s) and has been properly
configured, it is ready to accept the
PHD6 gas detector.
The PHD6 must be
turned on when it is
placed in the dock.
Note: If the PHD6 has
just been turned on,
wait at least three
minutes after the
conclusion of the
startup sequence to
allow the readings to
fully stabilize before
placing the instrument
in the dock.
Note: If a sensor has
just been replaced, it must be allowed
to stabilize before placing the
instrument in the dock. See the
sensor replacement section of the
PHD6 Reference Manual
for further information
on sensor stabilization.
If you plan to test the
instrument with the pump
attached, remove the
black spring clip from the
front of the dock prior to
inserting the PHD6 into
the dock.
Once the dock recognizes
the instrument, it will show
the following screens.
Once the connection is made, the IQ
Express will display “Connected”,
followed by the instrument type, software
version and recognized sensors.
If the connection is lost the following
screen will be shown:
If the instrument turns off prior to the
completion of the connection, Failed
will be shown: