background image

PP600F/602F LED lighting controllers - User Manual

Note:

 A PP600 is shown in these illustrations. The same power

connections apply for a PP600F or PP602F.

22

Summary of Contents for PP600F LED

Page 1: ...PP600F 602F LED Lighting Controllers Issue 012 User manual www gardasoft com...

Page 2: ...this equipment cannot cause any dangerous situation or significant financial loss to occur Deliberate acts of endangerment and vandalism are not covered by this document and must be considered by the...

Page 3: ...17 7 2 Total heat output 18 8 Connections 19 8 1 PP602F CCS connectors 20 8 2 Power supply 21 8 2 1 Single supply applications 21 8 2 2 Split supply applications 23 9 Operation 24 9 1 Startup 24 9 2...

Page 4: ...PP600F 602F LED lighting controllers User Manual 12 Error codes 39...

Page 5: ...ed to as PP600F Read Section 2 Safety or Section 3 Sicherheit or Section 4 S curit and Section 10 Reference information and check the PP600F fulfils your requirements Connect the PP600F up to a supply...

Page 6: ...to connect the outputs correctly and protect the output wiring and load from any short circuits When switched off energy remains stored in the PP600F for about 15 seconds The PP600F does not have com...

Page 7: ...roller must be externally fused to 4A using a slow blow fuse T4AH 50V The installer must provide a clearly marked nearby and easily accessible switch as part of the installation to allow the controlle...

Page 8: ...hentliche Ber hrung durch das Personal ausgeschlossen ist und stellen Sie sicher dass Luft frei um das Ger t zirkulieren kann 3 2 Elektrik Das PP600F erzeugt Impulse mit hoher Energie Achten Sie darau...

Page 9: ...sofern angebracht gesch tzt sein Die Verkabelung der Stromversorgung zur Steuerung muss f r min destens 4 A bemessen sein Die Verkabelung vom Kanalausgang zur Last muss h her als der maxi male Kanala...

Page 10: ...ts beeintr chtigt werden 3 4 Installationsanleitung Haftungsausschluss Diese Informationen dienen nur zur Orientierung Installationstechniker m ssen ihre eigene spezifische Risikobewertung f r die je...

Page 11: ...e pas de le toucher par accident et veillez ce que l air circule librement autour de l unit 4 2 lectricit Le PP600F produit des impulsions d nergie lev es Veillez bien raccorder les sorties et ce que...

Page 12: ...rtie de la cha ne la charge doit avoir une capacit sup rieure au courant de sortie maximal de la cha ne Si le contr leur est mal r gl ou en cas de panne l nergie fournie par l alimentation au contr le...

Page 13: ...ement titre indicatif Les installateurs doivent effectuer leur propre valuation des risques pour chaque installation M me si Gardasoft Vision Ltd a pr par minutieusement ces conseils Gardasoft Vision...

Page 14: ...a constant current supply as small variations in voltage can cause large variations in light output Currents can be specified in 0 25mA steps below 1A and in 1mA steps above 1A Three modes of operatio...

Page 15: ...into the enclosure by more than 6mm The enclosure of the PP600F is used to dissipate power in the form of heat For this reason the material to which the unit is attached must be suitable preferably me...

Page 16: ...PP600F together Chose a PSU that limits its output current by design by setting the current limit on the supply if this feature exists or use fuses Remember to de rate the fuse if mounted in an enclo...

Page 17: ...1 VL Voltage across the light LD1 and LD1 for output 1 This is usually easy to calculate as the voltage across the light is normally the voltage rating of the light as declared in its specification Al...

Page 18: ...r than 10W no heatsinking is required If the heat output is between 10W and 24W the PP600F needs to be mounted on a solid piece of metal to dissipate the heat Above 24W it is necessary to have a large...

Page 19: ...power connector has the following connections Screw terminal ID Function GND1 Channel 1 Power supply VS1 Power supply LD1 Load LD1 Load VS Digital power supply GND Digital power supply GND LD2 Channel...

Page 20: ...rs are not labelled on the case Refer to the table and the diagram below for the layout Connector Channel Use SKT9 1 CCS 12V lights SKT10 1 CCS 24V and HLV lights SKT12 2 CCS 12V lights SKT11 2 CCS 24...

Page 21: ...Section 8 2 1 Single supply applications It is possible to run the two channels and digital system from two or three separate supplies as shown in Section 8 2 2 Split supply applications though this...

Page 22: ...PP600F 602F LED lighting controllers User Manual Note A PP600 is shown in these illustrations The same power connections apply for a PP600F or PP602F 22...

Page 23: ...supply applications It is possible to run the two channels and digital system from two or three separate supplies as shown below though this will rarely be necessary Note It is essential that the grou...

Page 24: ...output CON The output current is fixed Selected continuous output SE 1 SE 2 SE12 One or both inputs can be used to select an output current For each state of the inputs the output current is fixed Pul...

Page 25: ...s received on one of the digital inputs the PP600F will wait for a delay period and then pulse the output The delay period pulse width and pulse current are all configurable Mode PU 1 triggers from IN...

Page 26: ...ry default the maximum rating is set to 4A 9 3 5 Setting numeric values When the PP600F displays numeric values for the user to change the right hand digit flashes to indicate that the p and q buttons...

Page 27: ...q buttons to select the required current in Amps Note Refer to Section 9 3 5 Setting numeric values Press SELECT dONE is displayed for one second The PP600F sets the output current 9 4 2 Setting pulse...

Page 28: ...displayed Wait for one second Use the p and q buttons to select the required current in Amps Note Refer to Section 9 3 5 Setting numeric val ues Press SELECT DEL is displayed Wait one second Use the...

Page 29: ...c values Press SELECT dONE is displayed for one second The PP600F is ready to pulse the output on a trigger Note 1 When setting up a PP600F controller pulsing on output channel 1 can only be triggered...

Page 30: ...the PP600F to be set up It is entered by holding the RANGE button down for one second In order for the whole range of timing values to be entered it is necessary to specify whether the delay is in se...

Page 31: ...s to select which output channel to set up CH1 or CH2 Press SELECT rAT is displayed Wait one second Use the p q and RANGE buttons to select the maximum current rating of the LED unit Press SELECT Use...

Page 32: ...the current in Amps a or C 3 to enter the current in milliamps b Press SELECT a b Use the p and q buttons to select the required current in Amps a or milliamps b Note Refer to Section 9 3 5 Setting n...

Page 33: ...RANGE for one second Use the p and q buttons to select which output channel to set up CH1 or CH2 Press SELECT rAT is displayed Wait one second Use the p q and RANGE buttons to select the maximum curre...

Page 34: ...nt in milliamps b Press SELECT a b Use the p and q buttons to select the required current in amps a or milliamps b Note Refer to Section 9 3 5 Setting numeric values Press SELECT dEL is displayed Wait...

Page 35: ...displayed Wait one second Use the p and q buttons to select the required pulse width in milliseconds Note Refer to Section 9 3 5 Setting numeric values Press SELECT dONE is displayed for one second T...

Page 36: ...able level IPH Input disable level IPL 24 IPH 4 5V IPL 1V This is the voltage applied between the anode and cathode of each optically isolated input Output current per channel 4A continuous or10A puls...

Page 37: ...ys are approximately 8 s shorter for a 24V trigger input The delays for channel 2 are greater All timings are given in microseconds Programmed Actual timing Output current 100mA Actual timing Output c...

Page 38: ...of 18V to 24V As with the standard range of controllers a trigger input of only 5V will increase the delay by approximately 8 s After a trigger occurs the PP600F controller halts all other operations...

Page 39: ...C E16 When the unit has powered up the non volatile memory was corrupt and has been cleared Set up the unit again E25 E26 The PP600F is calibrated during manufacture This stored calibration has becom...

Page 40: ...PP600F 602F LED lighting controllers User Manual Code Meaning E17 E40 E41 E64 E65 E67 E74 An internal software error has occurred Make a note of the number and contact your distributor 40...

Page 41: ...PP600F 602F LED lighting controllers User Manual This page is left blank for your notes 41...

Page 42: ...May 2017 Copyright 2017 Gardasoft Vision Ltd Gardasoft Vision Ltd Trinity Court Buckingway Business Park Cambridge CB24 4UQ UK tel 44 1954 234970 fax 44 1954 231567 Gardasoft LLC Oak Ridge Road Weare...

Reviews: