Panel Indicators and Connectors
45
Chapter 6:
Panel Indicators and Connectors
This chapter fully describes all the front and rear panel input and output devices on the Interface 5000. USB
commands control all Interface 5000 functions, so this information is only needed for troubleshooting and
special functions.
Front Panel
Power Button
The
Power
button is on the far right side of the Interface 5000 Front Panel. It is a push-push switch: push once
to turn the instrument on and push again to turn the instrument off.
The switch’s button latches
: the button is
closer to the front panel when the instrument is on than when it is off.
Normally the
PWR
LED illuminates when the Interface 5000 is powered on; see the
PWR
LED description
below.
Normally, the DC power is connected before the
Power
button is turned ON. However, no damage occurs if
this switch is already in the ON position when the cable is connected to
Power In
, or when the AC power input
is connected to the external power supply.
Power Switch
The PWR LED
The
PWR
LED is on the lower right of the Interface 5000 front panel. It normally glows a continuous blue when
the Interface 5000 is turned on and has passed some simple power-on tests.
Caution:
Avoid touching the
Power
switch button during normal instrument operation.
Data are lost if the instrument is accidentally powered off during an experiment.
The Interface 5000 uses three DIN connectors for connections to external devices. The three
DIN connectors all have different numbers of pins and physical sizes, so the cables used in them cannot be
interchanged or improperly connected.
Summary of Contents for Interface 5000
Page 2: ...2 ...
Page 18: ...Introduction 18 ...
Page 30: ...Installation 30 ...
Page 36: ...Calibration 36 ...
Page 64: ...Appendix A Interface 5000 Specifications 64 ...
Page 72: ...Appendix D Power LED Blink Codes and Error Messages 72 ...
Page 74: ...Appendix E CE Certificate Declaration of Conformity 74 Low Voltage Certificate of Conformance ...
Page 75: ...Appendix E CE Certificate Declaration of Conformity 75 RFI Certificate of Conformance ...
Page 76: ...Appendix E CE Certificate Declaration of Conformity 76 ...
Page 80: ...Appendix F Heat in Interface 5000 Multichannel Systems Interface Power Hub Systems 80 ...
Page 85: ......