Before dismantling Rediair to perform
filter maintenance/replacement,
a local risk assessment should
consider the appropriate PPE.
Always unplug the Rediair before
dismantling device to perform filter
maintenance/replacement.
WARNING
WARNING
Remove the external covers
and the pre-filters (both sides).
µg/m
3
PM2.5
1h 2h 4h 8h 12h
AUTO
POWER
MODE
SPEED
Switch off device and disconnect
the power supply.
Cleaning a pre-filter
• The surface of the pre-filter can be
vacuumed with a HEPA filter if available.
• The pre-filter can also be rinsed with
clean water but must be completely
dry before being used again as any
remaining moisture could potentially
cause damage to the HEPA 14 filter.
Cleaning and changing times
• Pre-filters are to be cleaned twice
a month and replaced twice a year.
• Main filters are to be
replaced twice a year.
• The ioniser is located within the
device and can’t be replaced.
• The particle sensor should be
cleaned every 2 months.
µg/m
3
PM2.5
1h 2h 4h 8h 12h
AUTO
FILTER
TIMER
LOCK
POWER
MODE
SPEED
FILTER
TIMER
LOCK
When filter indicator (3.6 on control
panel)
flashes red intermittently
,
it is time to clean the pre-filters.