11
External surface cleaning
Before performing any filter change,
clean/decontaminate the external
surface of the Rediair device. When
cleaning the exterior, remember to:
• Clean from top to bottom and
from clean to dirty areas.
• Wipe in an ‘S’ shape to avoid
recontamination of areas
you’ve already cleaned.
• If using a disinfectant, ensure that
recommended contact times are reached.
Filter changing and cleaning
Changing the main Rediair filter is
displayed by the filter indicator button
located on the digital display.
Pre-filters should be cleaned
regularly, at least once a month.
When the filter indicator changes to
red after 3000 hours (approximately
twice a year), you should replace the
main Rediair filter. After replacing the
Rediair main filter, reset the filter.
DISPOSAL, CLEANING AND MAINTENANCE
WIPE IN AN ‘S’
SHAPED PAT TERN
ENSURE CORREC T
CONTAC T TIME
WORK FROM TOP
TO BOT TOM
ONE WIPE,
ONE SURFACE
WIPE FROM CLEAN
TO DIRT Y
Clinell Universal
combined cleaning
& disinfecting wipes
are compatible with
Rediair. Additionally, for
a full list of compatible
agents, please refer
to the appendix.