4.3.4 Manual Zero
The
Manual Zero
button starts the zero calibration sequence of the analyzer which
incorporates the checking and adjusting of the instrument zero (baseline). No span correction
is required under normal circumstances on an installed analyzer. The calibration sequence is
identical whether trigger manually or automatically.
When triggered, the calibration sequence proceeds in the following order:
a) The
Mode
relay outputs changes to
Calibrate
and the
Control
relay changes to Manual.
b) The zero gas solenoid is energized and zero gas is introduced into the measurement cell
and is allowed to purge the measurement cell for the number of seconds specified by the
Zero Purge Interval
configuration parameter (Section 4.8.xx).
c) At the conclusion of the
Zero Purge Interval
, the computer averages the last ten (10)
spectrometer scans and determines a new reference scan vector (captures the new zero
value). At this time the zero gas solenoid is de-energized to allow sample gas to flood
the cell.
d) The analyzer waits for the number of seconds specified by the
Zero Hold Interval
parameter of the configuration data (Section 4.8)
e) During this period, process sample gas is drawn into the measurement cell and the
displayed concentration values will rise from zero to the measured values. The
Zero Hold
Interval
allows the analyzer concentration outputs to stabilize at on-line values before the
outputs are updated.
f)
Mode
and
Control
relay outputs signal change to Normal at the expiration of the
Zero
Hold Interval
(barring a pre-existing fault condition).
4.3.5 Back Purge
The default condition of the analyzer is to have sample gas flowing through the sample cell.
When
Back Purge
is selected, the analyzer sampling system is continuously back purged
with zero gas. This is useful as a precursor to maintenance activities or when long term
purging is required.
4.3.6 Graphic Display
The
Graphic
display of the H
2
S
concentration is updated once per second. The time and
intensity scales can be set on the
Config
panel (Section 4.8).
Summary of Contents for 962-AGX
Page 43: ...Revision 1 43 9 10 2019 Figure 4 13 Network Sub Panel...
Page 86: ...Revision 1 86 9 10 2019 Figure 9 1 Customer Connections...
Page 87: ...Revision 1 87 9 10 2019 Figure 9 2 Analyzer Components...
Page 88: ...Revision 1 88 9 10 2019 Figure 9 3 Inputs and Outputs...
Page 89: ...Revision 1 89 9 10 2019 Figure 9 4 AC Wiring Overview...
Page 90: ...Revision 1 90 9 10 2019 Figure 9 5 DC Wiring Overview...