Performance Characteristics
The performance characteristics for MR‐100™Plus Manual Resuscitators
will vary from user to user depending on a variety of factors: ambient
temperature, patient lung compliance, ventilator frequency & size of
operator's hands. The following data has been derived from MR‐100™ Plus
Manual Resuscitators in accordance with ISO standards.
Frequency range
-18
o
C/ 0
o
F
Cycle rate
22
o
C/72
o
F
Cycle rate
50
o
C/ 122
o
F
Cycle rate
Adult
38
45
45
Child A
92
105
105
Child B
72
78
76
Infant A
95
95
95
Infant B
60
60
60
These results were obtained under the following conditions:
Adult: VT- 600ml, Compliance -0 02L/cmH
2
O, Resistance - 20cmH
2
O/L/s
Child A: VT- 70ml, Compliance -0 01L/cmH
2
O, Resistance- 20cmH
2
O/L/s
Child B: VT- 300ml, Compliance -0.02 L/cmH
2
O, Resistance- 20cmH
2
O/L/s
Infant A: VT- 20ml, Compliance -0.001 L/cmH
2
O, Resistance- 400cmH
2
O/L/s
Infant B: VT-70ml, Compliance -0 01 L/cmH
2
O, Resistance- 20cmH
2
O/L/s
Specifications
Storage temp:
- 40'C/-10'F to 6 0 ' C / 140'F
Operating temp:
- 1 8 ' C / O'F to 50'C/ 1 22'F
Materials:
Silicon rubber:
Mask, Duckbill valve, Mushroom
valve, Relief valve seal, Bag body,
Oxygen reservoir bag, Oxygen
reservoir bag band & plug.
Polysulfone:
Non-rebreathing valve housing,
Pressure relief valve housing, A l l - in
one intake valve housing, Pressure
relief valve stem & cap.
Poly-vinyl-chloride: Oxygen tubing
Stainless steel:
Pressure relief valve
spring.
Aluminum:
Pressure relief n a m e plate.
EPDM:
O-ring
Connections:
Patient port
Silicon bag neck
Oxygen gas inlet
Intake valve port
Reservoir bag port
25 mmlD
15mmlD/22mmOD
24mmlD
6 m m O D
2 5 m m O D