Performance Characteristics
Range of delivery pressure
Adult: Unrestricted or 60±10 cmH
2
O
Child and Infant: 40±5 cmH
2
0 maximum (can be overridden by operator)
Stroke volume range (typical values)
Adult
Using one hand
800ml
Using two hands
1350ml
Child
350ml
Infant
160ml
Oxygen concentration
Adult
(Values In parentheses are without an oxygen reservoir 0
2
flow In Tidal vol. (ml) x ventilation rate)
L/min
600X12
800X20
750X12
750X20
1000X20 1000X20
05
82(33)
56(33)
65(33)
50(30)
55(30)
45(30)
10
99(37)
60(37)
99(37)
99(37)
66(37)
62(37)
15
99(45)
99(45)
99(45)
99(45)
99(45)
66(45)
Conditions:
compliance ‐ 0.02L/cmH
2
O, Resistence‐20 cmH
2
O/L
/s.
l:E ratio: 1 :2
Child
(Values In parentheses are without an oxygen reservoir 0
2
flow In Tidal vol. (ml) x ventilation rate)
L/min
70x20
70X30
200X20
200X30
300X20
300X30
02
99(65)
95(65)
56(37)
50(37)
45(34)
40(32)
05
99(86)
99(83)
99(44)
96(43)
84(37)
88(36)
10
99(93)
99(90)
99(46)
99(45)
99(45)
99(43)
Conditions:
compliance ‐ 0.01L/cmH
2
O, Resistance ‐ 200 cmH
2
O/L
/s.
l:E ratio: 1 :2
Infant
(Values In parentheses are without an oxygen reservoir 0
2
flow In Tidal vol. (ml) x ventilation rate)
L/min
20X30
20X60
40X30
40X60
70X30
70X60
02
99(72)
99(70)
99(70)
99(58)
97(55)
85(50)
05
99(77)
99(75)
99(74)
99(72)
99(65)
85(62)
10
99(94)
99(90)
99(90)
99(80)
99(79)
99(71)
Conditions:
VT=20ml, Compliance ‐ 0.001L/cmH
2
O, Resistance ‐ 400 cmH
2
O/L
/s.
l:E ratio: 1 :1
VT=40 & 70ml, Compliance ‐ 0.01L/cmH
2
O, Resistance ‐ 20 cmH
2
O/L
/s.
l:E ratio: 1 :2