background image

2.

 

Compress and hold the bag. The housing duckbill valve inside the non‐

rebreathing valve should open and test lung should be filled. If not, 

check      the connection between the Resuscitator and the test lung, and 

check that the non‐rebreathing valve is correctly assembled. 

3.

 

Release the bag. The duckbill valve should close and as the test lung     
deflates, gas should flow through the expiratory ports of the non‐

rebreathing valve. If not, check if the non‐rebreathing valve is correctly 

assembled. 

4.

 

Ventilate the test lung for a minimum of ten cycles to ensure that the 

Resuscitator is functioning correctly. Inspiration must occur when       

the silicone bag is compressed and exhalation should be in place, when 

the bag is released. If not, check on the non‐rebreathing valve is 

correctly assembled. 

 

To check the function of the pressure relief valve (infant and Child 

Resuscitators) 

Connect a 0‐100 cmH

2

0 manometer to the patient outlet of the non‐

rebreathing valve. Compress the bag. When the pressure relief valve 

activates, the manometer should read 35‐45cmH

2

0. If not, check that the 

non‐rebreathing valve is correctly assembled and does not leak. If the 

pressure relief valve fails a further test, it must be replaced. Do not attempt 

to repair the pressure relief valve. 

 

Oxygen Reservoir/Intake Valve assemblies

 

1.

 

Attach the reservoir to the intake valve inlet. 

2.

 

Inflate the reservoir and block the reservoir port. 

3.

 

Compress the reservoir bag. Gas should escape through the safety outlet 

valve on the reservoir valve. If not, check if the reservoir valve is 

correctly assembled. 

4.

 

Cycle the Resuscitator through several ventilations. The safety inlet valve 

on the reservoir valve should open during each refill to allow room air to 

enter the silicone bag. If not, check if the safety inlet valve is correctly 

assembled. 

 

Overall Resuscitator function 

1.

 

Fully assemble the Resuscitator (non‐rebreathing valve, silicone bag and       

oxygen reservoir). Connect the Resuscitator to a supplemental gas source 

and connect a test lung to the patient outlet on the non‐rebreathing 

valve. 

2.

 

Set the supplemental gas flow to 15LPM for the adult and the child 

models; and 10 LPM for the infant model. 

3.

 

Cycle the Resuscitator through several ventilations. The test lung should 

inflate during inspiration and deflate during exhalation. Check for 

leakage at all joints and connections. Ensure that the Resuscitator ref ills 

promptly and properly and that all valves are operating correctly If not, 

repeat the tests above to find where the problem lies. 

 

 

Accessories 

1.

 

Masks 

  masks from infant to adult (6 sizes). 

2.

 

Peep valve 

Durable 2‐10 adjustable peep valve.   

Durable 5‐20 adjustable peep valve.   

Diverter for peep valve 

Summary of Contents for 2038

Page 1: ...REF SAP REF Description 2038 AR0051 MR 100 Plus Intake Valve Kit ...

Page 2: ...uld read and understand the contents of this manual and demonstrate proficiency in the assembly disassembly and use of this device prior to use Warning And Cautions WARNING Do not use the MR 100 Plus Manual Resuscitator In toxic atmospheres WARNING Remove the oxygen reservoir and reservoir valve If supplemental oxygen Is not being administered Failure to do so will affect the refill rate and maxim...

Page 3: ...ent side of the duckbill valve F or when the operator releases pressure from the silicone bag 8 The valve lifts directing the patient s expiratory gases through the expiration ports G of the non rebreathing valve see figure 3 Figure 3 lnspiration The silicone bag refills for the next breath during patient exhalation Negative pressure within the bag caused by the expansion of the compressed bag ope...

Page 4: ...will stay in the overridden position Lock Arrow forward to the NRV direction as Figure 5 1 Figure5 1 Figure 5 Overriding the Pressure Relieve Valve Operating Instruction Note lf the patient is intubated or has a tracheostomy the resuscitator can be connected by removing the mask and attaching the non rebreathing valve outlet directly to the connector of the endotracheal or tracheostomy tube 1 Posi...

Page 5: ...g valve as follows Squeeze the silicone bag to deliver several sharp breaths through the non rebreathing valve to expel the contamination If the contamination does not clear continue ventilation with another resuscitator or use mouth to mouth mask techniques Disassemble the non rebreathing valve and rinse it with water Reassemble and disinfect Once the procedure is completed clean disinfect and te...

Page 6: ...es are without an oxygen reservoir 02 flow In Tidal vol ml x ventilation rate L min 70x20 70X30 200X20 200X30 300X20 300X30 02 99 65 95 65 56 37 50 37 45 34 40 32 05 99 86 99 83 99 44 96 43 84 37 88 36 10 99 93 99 90 99 46 99 45 99 45 99 43 Conditions compliance 0 01L cmH2O Resistance 200 cmH2O L s l E ratio 1 2 Infant Values In parentheses are without an oxygen reservoir 02 flow In Tidal vol ml x...

Page 7: ...ance 0 01L cmH2O Resistance 20cmH2O L s Child B VT 300ml Compliance 0 02 L cmH2O Resistance 20cmH2O L s Infant A VT 20ml Compliance 0 001 L cmH2O Resistance 400cmH2O L s Infant B VT 70ml Compliance 0 01 L cmH2O Resistance 20cmH2O L s Specifications Storage temp 40 C 10 F to 60 C 140 F Operating temp 18 C O F to 50 C 122 F Materials Silicon rubber Mask Duckbill valve Mushroom valve Relief valve sea...

Page 8: ... rebreathing valve outlet 3 Release the bag The bag should expand immediately and refill If not check the intake valve at the base of the silicone bag is correctly assembled or not 4 While keeping the non rebreathing valve outlet blocked compress the bag again The bag should not be compressed easily If this occurs check that you are blocking the valve sufficiently and that the intake valve at the ...

Page 9: ...ressure relief valve Oxygen Reservoir Intake Valve assemblies 1 Attach the reservoir to the intake valve inlet 2 Inflate the reservoir and block the reservoir port 3 Compress the reservoir bag Gas should escape through the safety outlet valve on the reservoir valve If not check if the reservoir valve is correctly assembled 4 Cycle the Resuscitator through several ventilations The safety inlet valv...

Page 10: ...nd Storage Wall mount holder for resuscitators Silicone hanger for resuscitators Storage case Order information REF SAP REF Description 2038 AR0051 MR 100 Plus Intake Valve Kit 2150 AR0056 MR 100 Plus Resuscitator set adult w Pop 2151 AR0057 MR 100 Plus Resuscitator child 2152 AR0058 MR 100 Plus Resuscitator infant 2163 AR0063 MR 100 Plus Resuscitator adult ...

Reviews: