EVO EXPERT Manual
20/105
Upon entering this screen, the spindle field is highlighted in light blue background. Using the ‘
’ and
‘
’ keys you can change the Spindle.
The list of possible spindles to use depends on the model of your viscometer (L, R or H). Thus, in
tables 8 through 22 (page 65 and on) you can see the different spindles available for each model.
Note:
It is recommended to use the following combinations of Spindle and standard
oil depending on the viscometer model:
Model L:
TL5 Spindle (LCP adapter) and RT-50 (50 cP) standard
Model R:
TR8 Spindle (LCP adapter) and RT-500 (500 cP) standard
Model H:
TR8 Spindle (LCP adapter) and RT-500 (500 cP) standard
Once you’ve selected your spindle press ‘ENTER’ and the ‘VISCOSITY’ field will be highlighted. Press
‘ENTER’ again and the following screen will appear:
Use the numerical keyboard to introduce the value of the viscosity of the standard oil used for
calibration (the standard oils provided by Fungilab provide viscosity tables according to different
working temperatures). There is a field for entire numbers and other one for the decimal figures.
Once introduced the density value, press ‘ENTER’ with the ‘SAVE’ option highlighted to confirm the
modification. Next, the following screen will appear:
Remove the Spindle if it is connected to the viscometer and press ‘ENTER’. Then, the viscometer
will perform the Offset calibration, showing the following screen:
The screen shows the progression of this step of the calibration with a status bar. Once the Offset
calibration is completed, the following screen will be on: