EVO EXPERT Manual
54/105
Once the data download is completed the viscometer will return to the main menu. However, if the
connection with the WiFi network or the FTP server is not properly established the instrument will
show the following screen:
Examine the WiFi configuration parameters and execute the download again. The configuration of
the instrument to be connected to a WiFi network will be explained in the Appendix A, ‘Wireless
Datalogger configuration’.
Note:
For some particular WiFi network conditions, such as temporary overloading, some
loss of data may result during the download. This would lead to downloaded files with
missed recordings. This issue can be solved downloading again the experiment stored in
the memory of the viscometer, at any time before saving a new experiment.
8.6.3 Information
If you select the ‘INFO’ option, you will be brought to the following screen:
The geographic area can be chosen here use the ‘
’ and ‘
’ keys. Press the ‘ENTER’ key and it will