63
13. Maintenance and Inspection
This section explains the daily checks and periodic checks of the
LX-8300M. To ensure safety, reliability, and high performance, a “Daily Check”
and “Periodic Check” must be performed. We are not liable for any accident
arising from lack of maintenance. A full technical description of the LX-8300M
is available from your local Fukuda Denshi representative.
CAUTION
Do not open the housing.
Do not allow excessive moisture or cleaning agents into the connectors
or inside the equipment.
■
Daily Check
Perform daily checks using the “Daily Check List” on the next page.
■
Periodic Check
Periodic check of medical electronic equipment is mandatory to prevent
failures and accidents, and to ensure safety and reliability.
Periodic maintenance may be performed by the medical institution or by a third
party by concluding a “Maintenance Contract”.
For more details, contact your local Fukuda Denshi service representative.
Perform Periodic check using the “Periodic Check List”.
The periodic check should be performed once a year.
If there is an item with “Fail” judgement, the overall judgement will be “Fail”.
Make sure to take countermeasures for the “Fail” item.
Use the equipment only if the judgements of all the items are “Pass”.
■
Periodic Replacement Parts
The “Battery Compartment Lid (Waterproof)” is the only periodic replacement
part.
To ensure the reliability of waterproof (IPX8) performance of the LX-8300M,
replace it once a year.
Contact your local Fukuda Denshi service representative for replacement.
The reliability of water resistance (IPX8) performance will not be ensured
without yearly replacement.
CAUTION
The periodic replacement parts must be replaced at specified period.
Summary of Contents for LX-8000 Series
Page 1: ......
Page 2: ......
Page 22: ...xx ...
Page 26: ...1 General Description 2 ...
Page 34: ...3 Preparation 10 ...
Page 40: ...5 Respiration Monitoring 16 ...
Page 54: ...6 SpO2 Monitoring 30 ...
Page 74: ...9 Other Setting Items 50 ...
Page 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Page 84: ...11 Troubleshooting 60 ...
Page 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Page 102: ...14 Standard and Optional Accessories 78 ...
Page 108: ...15 Specification 84 ...
Page 109: ......
Page 110: ......