11. Troubleshooting
55
ECG
Message/Icon Cause
Solution
The displayed
character string
indicates the
detached electrode
position.
For details, refer to
“Details of the
“Electrode”
Message”.
Electrode is off.
Check the electrode
condition.
Lead cable is off.
Check the connection
between the lead cable and
this equipment.
Check the connection
between the lead cable and
the electrode.
Faulty Lead cable.
Replace the ECG cable with a
new one.
Electrode is peeling off.
Replace the electrode with a
new one.
Polarization potential of
the electrode is too high.
Replace the electrode with a
new one.
Situation Cause
Solution
ECG waveform
contains noise.
Electrode gel is dry
Replace the electrode with a
new one.
Electrode is peeling off.
Electric blanket is used.
Cover the electric blanket
with a shield cover.
AC filter setting of the
receiving monitor is OFF.
Set the AC filter up as ON.
Respiration waveform
cannot be measured.
Electrode gel is dry
Replace the electrode with a
new one.
Electrode is peeling off.
The positions of the
electrodes are improper.
Attach the electrodes where
the respiration waveform can
be measured appropriately.
Summary of Contents for LX-8000 Series
Page 1: ......
Page 2: ......
Page 22: ...xx ...
Page 26: ...1 General Description 2 ...
Page 34: ...3 Preparation 10 ...
Page 40: ...5 Respiration Monitoring 16 ...
Page 54: ...6 SpO2 Monitoring 30 ...
Page 74: ...9 Other Setting Items 50 ...
Page 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Page 84: ...11 Troubleshooting 60 ...
Page 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Page 102: ...14 Standard and Optional Accessories 78 ...
Page 108: ...15 Specification 84 ...
Page 109: ......
Page 110: ......