17
6. SpO
2
Monitoring
When the transmitter is used without the ECG lead cable, it will measure only
SpO
2
.
WARNING
When the SpO
2
probe (sensor) is in a connector-off condition, the SpO
2
alarm will not be generated on the receiving monitor regardless of the
SpO
2
measurement ON/OFF status. Make sure that the SpO
2
probe
(sensor) is securely connected. If the SpO
2
waveform/numeric data is not
displayed, check the patient’s condition and pay attention not to miss the
connector-off condition.
If the display is OFF, SpO
2
measurement will not start even when the SpO
2
probe is connected. In such case, turn ON the display. SpO
2
measurement
will automatically start regardless of the ON/OFF status of SpO
2
measurement.
CAUTION
When using the transmitter with only the SpO
2
sensor cable, ECG and
respiration measurements on the receiving monitor shall be turned off to
prevent an erroneous alarm.
The pulse wave and level meter are normalized for SpO
2
measurement.
It does not indicate perfused blood volume. Check proper probe
attachment by observing the pulse wave.
■
Preparation for Monitoring
Select the proper SpO
2
sensors depending on the purpose and application
sites defined in the sensors directions for use.
For details of the usable SpO
2
sensors, refer to “14. Standard and Optional
Accessories”.
WARNING
For SpO
2
monitoring, always use the sensor specified by Fukuda
Denshi. Also, check the probe before usage to make sure that it is the
specified probe. If unspecified probe is used, not only that the equipment
cannot deliver its maximum performance, the equipment may be
damaged and safety cannot be ensured.
As with all medical equipment, carefully route cables to reduce the
possibility of patient entanglement and strangulation.
Do not place this equipment or accessories in any position that might
cause it to fall on the patient.
Do not start or operate this equipment unless the setup was verified to
be correct.
Do not operate this equipment in magnetic resonance imaging (MRI)
environments.
Summary of Contents for LX-8000 Series
Page 1: ......
Page 2: ......
Page 22: ...xx ...
Page 26: ...1 General Description 2 ...
Page 34: ...3 Preparation 10 ...
Page 40: ...5 Respiration Monitoring 16 ...
Page 54: ...6 SpO2 Monitoring 30 ...
Page 74: ...9 Other Setting Items 50 ...
Page 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Page 84: ...11 Troubleshooting 60 ...
Page 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Page 102: ...14 Standard and Optional Accessories 78 ...
Page 108: ...15 Specification 84 ...
Page 109: ......
Page 110: ......