Chapter 1
1 - 6
Specifications list
Leads
Standard 12 leads
Sensitivity
1/4, 1/2, 1, 2cm/mV, or automatic
Polarization voltage
±
550mV or greater
Frequency response
0.05 to 150Hz
Time constant
3.2 sec
Common mode rejection
2mmp-p or lower (103dB or more)
Input impedance
50M ohms or greater
Input circuit current
5 x 10
-8
A or lower
Internal noise
20
µ
Vp-p or lower
Calibrated voltage
1mV
±
2% or lower
Sampling rate
8000/Sec/CH
Filter
HUM filter:
-20dB or less at 50/60Hz
Muscle filter: -3dB (-6dB/oct) at 25/35Hz
Drift filter:
-3dB or less at 0.25/0.5Hz
Recording system
Thermal array head
Recording speed
5, 10, 12.5, 25, 50mm/sec.
Recording resolution
Deflection axis: 8 dots/mm
Time axis:
500
μ
sec.
Recording channels
3CH, 3CH+rhythm lead, 6CH
Recording paper
Roll recording paper: OP-358TE (145mm x 30M)
Z-fold paper: OP-383TE (145mm x 100M)
Display
Type:
Liquid crystal display (LCD) with backlighting
Resolution: 320 dot (horizontal) x 240 dot (vertical)
Electro-
cardiography
Patient data
ID number, age, sex, height, weight, etc.
Basic measurement value
Heart rate, R-R time, P-R time, QRS time, QT time, QTc,
electrical axis, SV1, RV5(6)
Heart Rate measurement
20 to 300 BPM
Measurement
Operation
Accuracy
±
2 BPM
Interpretation and code
Approx. 110 types
Minnesota code
Approx. 130 types
Analysis operation
(When the
PC-7303 or
PC-7304 program
cartridge is
installed)
Grade judgment
4 types
Safety standard
IEC60601-1 (Class I, type CF; Internally powered
equipment, type CF)
Power supply
AC power supply: 100 to 240V AC (
±
10%), 50/60Hz,
Power consumption: 80VA
DC power supply: Ni-MH
battery
(T8HR4/3FAUC-5345),
Power consumption: 60W
Dimensions
315mm (W) x 255mm (D) x 76mm (H)
(12.4” x 10.0” x 2.99”)
General
Weight
Approx. 3.5kg (7.7 lbs.)
Temperature
+10 to +40 degrees Celsius (50 to 104 degrees
Fahrenheit)
Operating
environment
Humidity
25 to 95% (no condensation)
Atmospheric pressure
70 to 106kPa (700 to 1060mbar)
Storage
environment
Temperature
-10 to +50 degrees Celsius (14 to 122 degrees
Fahrenheit)
Humidity
10 to 95% (no condensation)
Atmospheric pressure
70 to 106 kPa (700 to 1060mbar)
Summary of Contents for FX-7302
Page 10: ...viii ...
Page 12: ......
Page 60: ...Chapter 2 2 40 ...
Page 62: ......
Page 86: ...Chapter 4 4 18 ...
Page 87: ...4 19 MAIN BOARD circuit diagram PCB 6956B Power connection table No 1 ...
Page 114: ...4 46 ...
Page 116: ......
Page 130: ......
Page 131: ...6 1 MAIN BOARD PCB 6956B component side ...
Page 132: ...6 2 MAIN BOARD PCB 6956B soldering side ...
Page 133: ...6 3 OPERATION BOARD PCB 6957B component side ...
Page 134: ...6 4 OPERATION BOARD PCB 6957B soldering side ...
Page 135: ...6 5 ECG AMP BOARD PCB 6731D component side ...
Page 136: ...6 6 ECG AMP BOARD PCB 6731D soldering side ...
Page 137: ...6 7 MARK SENSER BOARD PCB 6942 component side ...
Page 138: ...6 8 MAGAZIN UP BOARD PCB 6943A component side ...
Page 139: ...6 9 CONNECTION BOARD PCB 7028 component side ...
Page 140: ...6 10 ...
Page 142: ......
Page 151: ...7 9 Outline view 255 315 70 76 ...
Page 152: ...7 10 ...
Page 154: ......
Page 160: ...Chapter 8 8 6 ...
Page 161: ......