Chapter 3
3 - 2
Removing the MAIN BOARD and the ECG AMP unit
CAUTION
To prevent damage by static electricity, eliminate static electricity from
your body before touching the board.
1. Remove the enclosure cover.
For details on removing procedures, see page 3-1, “Removing the enclosure
cover.”
ECG amp
CF CARD
SERIAL
LAN
P5
Mounting screws
2. Remove the connector connected to the MAIN BOARD (P5).
3. Remove the mounting screws (four) first, and then the ECG AMP unit.
ECG amp
Mounting screw
P5