2 - 28
Checking the earth leakage current
(current that flows to protective earth resistance)
Measurement wiring diagram
Conversion plug
Used for single fault condition
FX-7302
MD
Leakage current measuring device
Power supply cord
Conversion plug
Measurement point A
Measurement
point B
Power outlet
Earth leakage current measurement points
Measurement point A
Earth wire of power supply cord from unit
Measurement point B
Earth terminal of power outlet
Earth leakage current measurement conditions
It is necessary to measure the earth leakage current under two conditions:
normal condition and single fault condition.
How to change the phase
You can switch the phase by inserting the conversion plug into the power
outlet upside down.
How to reproduce a single fault condition
To reproduce a single fault condition, insert one of the two inserts of the
conversion plug by cutting the other insert.
Acceptance standards
Normal condition
0.5 mA or less
Single fault condition
1.0 mA or less
Summary of Contents for FX-7302
Page 10: ...viii ...
Page 12: ......
Page 60: ...Chapter 2 2 40 ...
Page 62: ......
Page 86: ...Chapter 4 4 18 ...
Page 87: ...4 19 MAIN BOARD circuit diagram PCB 6956B Power connection table No 1 ...
Page 114: ...4 46 ...
Page 116: ......
Page 130: ......
Page 131: ...6 1 MAIN BOARD PCB 6956B component side ...
Page 132: ...6 2 MAIN BOARD PCB 6956B soldering side ...
Page 133: ...6 3 OPERATION BOARD PCB 6957B component side ...
Page 134: ...6 4 OPERATION BOARD PCB 6957B soldering side ...
Page 135: ...6 5 ECG AMP BOARD PCB 6731D component side ...
Page 136: ...6 6 ECG AMP BOARD PCB 6731D soldering side ...
Page 137: ...6 7 MARK SENSER BOARD PCB 6942 component side ...
Page 138: ...6 8 MAGAZIN UP BOARD PCB 6943A component side ...
Page 139: ...6 9 CONNECTION BOARD PCB 7028 component side ...
Page 140: ...6 10 ...
Page 142: ......
Page 151: ...7 9 Outline view 255 315 70 76 ...
Page 152: ...7 10 ...
Page 154: ......
Page 160: ...Chapter 8 8 6 ...
Page 161: ......