- 53 -
INZ-TN2ZSVS-E
(3) Feed the moisture interference gas using the
key.
Press the
key, and the cursor moves to the correc-
tion value.
(4) When the interference value stabilizes, enter correction
value using the
(UP) key,
(DOWN) key, and
the
(digit) key so that the interference value
becomes 0.
(5) Press the
key, and the data is registered.
The setting has now been completed.
(1) Move the cursor to <6. Interference> using the
key, and press the
key.
The screen switches as shown at right.
<Adjusting 2 components simultaneously by ALL>
(2) Move the cursor to ALL using the
key.
(3) Press the
key, and the moisture interference gas is
fed.
Press the
key, and the cursor is moved.
(4) When the interference value stabilizes, press the
key.
Automatic adjustment is carried out, and the compo-
nent selection screen appears again.
The setting has now been completed.
ENT
,
,
,
MODE
Setting complete
ESC
ESC
ENT
,
ENT
ENT
,
MEAS
Setting complete