- 23 -
INZ-TN2ZSVS-E
7) The cursor moves to
the reading.
Remove the pipe
connected to the
sample gas inlet.
8) Press the key, and the air is
automatically sucked from the
sample gas inlet. Check that the
flow rate reading of the flow
checker is kept
at approximately
0.5 L/min.
,
,
7) The cursor moves to
the reading.
Auto
9) After arbitrarily set time elapses,
the CH1 span calibration screen appears.
,
,
ESC
ESC
ENT
ENT
8) Connect N
2
gas to the calibration
gas inlet. Press the key and
check that the flow rate reading
of the flow checker is kept at
approximately
0.5 L/min.
ENT
ENT
ESC
ESC
ENT
ENT
CH1 span
Calibration screen
MODE
CH1 span
Calibration screen
MODE
MODE
CH1 span
Calibration screen
MODE
CH1 span
Calibration screen
MODE
<In the case of Air>
<In the case of Dry N
2
>
To forcibly perform zero calibration without waiting for the arbitrarily set time of calibration to elapse,
press the
key in 8).
How to forcibly perform zero calibration
• If O
2
analyzer is provided, even if “ON” is
selected for O
2
zero calibration, calibration is
not performed because O
2
is contained in the
zero gas.
Caution on operation
• Zero calibration is performed in batch for all the ranges of selected components.
• If “OFF” is selected for zero calibration of all the components, all zero calibrations are not performed.
• Zero calibration is performed at the time arbitrarily set. The time of completion of zero calibration
is displayed at the upper right-hand corner of the screen.
Caution on operation