14
FONIX 7000 Hearing Aid Analyzer
11.
PROBE:
Connects the main module to the remote module (used for
real-ear measurements).
KEYBD:
Not currently active.
AUX:
Not currently active.
The following safety symbols can be found on the back of the 7000 hearing aid
analyzer:
Type B Equipment. The probe microphone and insert ear-
phone are type B applied parts, according to IEC 60601-1.
Read the accompanying documents. Please read this manual
before operating the 7000 test system. A separate maintenance
manual exists for the 7000 test system. If you wish to obtain a
maintenance manual, please contact Frye Electronics, Inc. or
your Frye representative.
This symbol indicates that Frye Electronics, Inc. conforms to
the Medical Device Directive 93/42/EEC. Any attached video
monitor, external printer, or external computer should also
have a CE mark in order for the 7000 test system to remain
compliant.
Replace fuses only with the same type and rating.
1.5.3 Safety
Safety Classification for IEC 60601-1
Type of protection against electric shock: Class I
Degree of protection against electric shock: Type B
Protection against harmful ingress of water: Ordinary
Mode of operation: Continuous
The 7000 does not require sterilization or disinfection.
Warning:
This equipment is not suitable for use in the presence of flammable
anaesthetic mixture with air or with oxygen or nitrous oxide.