FONIX 7000 Hearing Aid Analyzer
Coupler Problems:
1. HA-2 Coupler (BTE) adapter tubing is missing, loose, or cracked
Replace with #13 thickwall tubing. Length: 0.6" (15mm).
2. Test microphone is difficult to get into coupler or the ear level adapter
does not easily seal to the other end of the coupler
Lubricate the black “O” ring with light petroleum type lubricant.
3. A bump or peak in the low frequency response curve
a. There may be a hearing aid vent leak. Be sure to Fun-Tak the vent.
b. There may be a coupler vent leak.
c. The #13 coupler tubing could be cracked or broken.
Probe Problems:
1. Probe Reference Mic does not Level.
a. Check the reference mic calibration. (See Appendix B)
b. Be sure the reference mic is properly plugged into the instrument.
c. Are all connections clean and tight?
d. Are the mic cables loose, broken, cut, worn or frayed?
e. Is the Leveling signal coming out of the speaker? If not, check cable and
connections.
f. Make sure the distance from the speaker to the ref. mic during leveling is
about 12” (max. 18”).
2. Cleaning probe tubes
DO NOT REUSE probe tubes. There is NO recommended cleaning proce-
dure. Germicidal solutions can leave a residue inside the tubing which can
cause test result errors. DO NOT cut off any portion of the tube.
Printer Problems:
1. Printer does not work
a. Check for a paper jam.
b. Press the FEED Button.
c. Make sure the print head lock-down lever is down.
2. Test results do not print on paper
Make sure you are using thermal paper. To check it, take a hard object, e.g.
a car key, and scratch the surface of the paper on both sides. If a black mark
appears it is thermal paper. If not, it is plain paper and will not work.