8
Cleaning
DANGER!
Mains voltage!
– Before beginning with cleaning work, disconnect
the mains plug and protect the device against being
unintentionally switched back on!
– Do not allow any liquids to flow into the device.
– Indicate cleaning work with warning signs.
– Put safety equipment back into operation after
cleaning work.
When cleaning the entire device, adhere to the guide-
lines of the Accident Prevention Regulation (BGV A3) -
especially when the device has been set up in a dusty
environment or when the grinding stock processed pro-
duces dust.
8.1 Grinding elements
NOTICE!
Cool grinding elements made of agate, sintered
corundum, zirconium oxide and silicon nitride slowly
and carefully.
Do not heat agate elements in a microwave under any
circumstances (heating is too fast).
They must never be exposed to thermal shocks as this
could cause irreparable damage to the parts
®
They
will burst apart like in an explosion.
n
Clean the grinding bowl and grinding balls each time after using
them: Clean them, e.g., under running water using a brush and a
commercially available cleaning agent.
n
Half fill the grinding bowl with grinding balls, sand and water, and
run with maximum speed for 2 to 3 minutes (correctly tensioned) in
the Planetary mono mill.
n
Cleaning with an ultrasonic cleaner is permitted.
n
For sterilisation in the heat cabinet, only heat the grinding elements
up to 100 °C.
Cleaning