2.6 Protective equipment
Protective equipment is to be used as intended and may
not be disabled or removed.
All protective equipment is to be regularly checked for
integrity and proper functioning.
For start-up, the hood (3) has to be closed.
The hood (3) is locked:
n
without mains connection
n
during operation
The hood (3) can only be opened, if the mill's drive is at
standstill.
2.6.1 Opening the hood without mains connection
NOTICE!
Use the emergency release only if a power failure or
damage to the device prevents normal opening. Only
open the emergency release gently and turn the trian-
gular key 180 degrees clockwise. Any other manner of
handling can damage the locking mechanism, rendering
the mill unusable.
CAUTION!
The emergency release must not be activated while the
machine is running! Disconnect the machine from the
mains before the emergency release. Failure to observe
this will render void the guarantee, and releases us
from liability for any resulting damage to the device or
personal injury.
Safety information and use