NOTICE!
Fritsch mills are speed controlled. The devices are
equipped for this with frequency converters. In order to
comply with the EMC directive, many measures must
be taken to prevent operational transient emissions.
The possible leakage currents resulting from filtering
measures can trigger a conventional residual current
circuit breaker in the mains line. This is no defect!
To prevent this, special residual current circuit breakers,
which are adapted for operation with frequency con-
verters, are commercially available.
Operation without a residual current switch is possible,
but must be done in accordance with the relevant regu-
lations.
4.5.1 Adjusting the mains voltage
4.5.1.1
Adjusting the mains voltage with the rotary switch (9)
CAUTION!
Only qualified personnel may change the voltage range
on the device!
CAUTION!
The voltage range may only be adjusted after the mains
has been disconnected. Disconnect the device from the
mains!
1.
Disconnect the device from the mains!
2.
The rotary switch (9) for adjusting the mains voltage is located on
the back side of the device. Rotate this switch to the required
voltage.
3.
Connect the device to the mains.
Installation
- 23 -