1.
Once the Final-Fill Sensor is replaced or the Unit is relocated and the empty sample bottles are
available, plug in the power cord. Turn treated water ON. Turn the power switch ON (Red switch on
the left side of the GranuFlo I Dissolution Unit). Verify the power light activates.
2.
Secure the Tank Lid and initiate the
RINSE CYCLE
according to instructions found in the Operator’s
Manual, section 6.0.
3.
Run a
DISSOLUTION CYCLE
according to instructions in section 7.0 of the Operator’s Manual. When
the ADD GRANULES light flashes, make sure water has reached the Mid-Level Sensor, and add the
eight (8) boxes of GranuFlo® Dry Acid product.
4.
Perform the
Specific Gravity
Test and record the results on the GranuFlo Batch Production Record
Form according to the Operator’s Manual, section 7.1
5.
Collect a sample of the final product(s) using the (2) sample bottles (P/N G83-535-02). The product
sample(s) will be analyzed in accordance to manufacturer’s product specifications. If only one
product code is being used, collect two (2) samples of the product to be analyzed.
6.
Place the sample bottles and the following completed forms into a shipping box:
•
Copy of the GranuFlo Batch Production Record Form (Operator’s Manual, pg 31 )
•
Batch Analysis Form (Operator’s Manual pg 53)
7.
Cut out and affix the pre-printed mailing label from the Batch Analysis Form to the box. It is the
responsibility of the RES to ship the samples to the “Ship To” address on the Batch Analysis Form.
Contact Fresenius Medical Care Laboratory: at (
972) 929-7291 for results.
8.
NOTICE: The composition of the first batch of GranuFlo® Dry Acid product must be tested by a
qualified testing laboratory to ensure that the resulting product meets the GranuFlo® Dry Acid
product specifications. If final solution did not meet final batch criteria for use, the batch of
concentrate must be discarded
(S
EE
S
ECTION
13:
DISPOSAL OF
C
ONCENTRATE
S
OLUTION
,
PG
NOTE:
MAKE
SURE
TO
FILL
THE
GRANUFLO
BATCH
PRODUCTION
RECORD,
PAGE
OF
THIS
MANUAL
WITH
THE
PRODUCT
INFORMATION.
WARNING:
DO
NOT
USE
ANY
BOX
OF
GRANUFLO®
DRY
CONCENTRATE
THAT
HAS
BEEN
OPENED
OR
TAMPERED
WITH.
IT
IS
IMPORTANT
THAT
THE
ENTIRE
CONTENTS
OF
EACH
BOX
ARE
EMPTIED
INTO
THE
DISSOLUTION
TANK.
WARNING!
THE
USE
OF
EYE
PROTECTION,
AND
GLOVES
IS
RECOMMENDED
WHEN
HANDLING
DRY
ACID
PRODUCT.
IF
CONTACT
WITH
EYES,
RINSE
IMMEDIATELY
FOR
15
MINUTES.
IF
CONTACT
WITH
SKIN,
FLUSH
WITH
PLENTY
OF
SOAP
AND
WATER.
SEE
MATERIAL
SAFETY
DATA
SHEETS
(MSDS)
FOR
THE
GRANUFLO®
DRY
ACID
PRODUCT
BEING
USED
FOR
FURTHER
PERSONAL
PROTECTIVE
EQUIPMENT
(PPE)
OR
EMERGENCY
REQUIREMENTS/INSTRUCTIONS.
Summary of Contents for GranuFlo 450385
Page 1: ... GranuFlo CONCENTRATE DISSOLUTION UNIT I Version 2 OPERATOR S MANUAL P N 450385 Rev H ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 38: ...32 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 40: ...34 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 42: ...36 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...38 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 48: ...42 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...P N 450385 Rev H 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 50: ...44 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 450385 Rev H 45 ...
Page 52: ...46 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 450385 Rev H 47 ...
Page 54: ...48 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 450385 Rev H 49 ...
Page 56: ...50 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 58: ...52 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 60: ...54 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 450385 Rev H 55 PAGE INTENTIONALLY LEFT BLANK ...