P/N 450385 Rev. H
23
S
ECTION
8:
M
AINTENANCE
The GranuFlo I Dissolution Unit is designed for ease of use and trouble-free operation. However, a minimal
amount of preventive maintenance is required in order to maintain the GranuFlo I Dissolution Unit in good
working condition and minimize the possibility of a system malfunction.
The recommended program for proper care of the GranuFlo I Dissolution Unit consists of four basic steps.
They are: (8.1) Regular visual inspection, (8.2) Cleaning, (8.3) Sodium Hypochlorite (Bleach) Disinfection, and
(8.4) Filter Maintenance.
8.1
V
ISUAL
I
NSPECTION
Visually inspect the GranuFlo I Dissolution Unit prior to mixing a batch. The Operator should look for
any defects which may inhibit the safe or proper operation of the Unit. Items such as damaged
hydraulic hoses or fittings, damaged electrical cables or connections, loose, missing or damaged
hardware, or process contamination should be corrected prior to the use of the GranuFlo I Dissolution
Unit.
Should the power cord or plug become cracked, frayed, or otherwise damaged, it should be replaced
immediately. Tag Unit “OUT of SERVICE” until repaired. Always remove the power cord from the power
source before attempting to service this device. Never unplug the Unit by pulling on the power cord.
8.2
C
LEANING
Clean the exterior surface of the GranuFlo I Dissolution Unit thoroughly after each batch of concentrate
is mixed. If necessary, a mild detergent solution may be used to clean the exterior surface. Care should
be taken not to contaminate the system interior. Wipe all spills immediately. Avoid spillage at the
control panel to minimize the possibility of electrical malfunction.
8.2.1.
P
REVENTIVE
M
AINTENANCE
Table 1: Routine Maintenance Schedule on pg. 30.
8.3
S
ODIUM
H
YPOCHLORITE
(R
EGULAR
B
LEACH
)
D
ISINFECTION
1.
To start the disinfection process, power OFF the GranuFlo I Dissolution Unit, remove the Filter
Housing and discard the Filter Element. Place the residual fluid from the Filter Housing in the
Residual Solution Bucket.
Reinstall the Filter Housing, but do not insert a Filter at this time. Connect
the Transfer Nozzle to the end of the Dissolution Unit’s Transfer Hose, and make sure the valve on
the Transfer Nozzle is in the CLOSED position. In addition, loosen the top Access Lid, but leave it in
place. Ensure you have the appropriate personal protective equipment donned.
WARNING!
T
O PREVENT ACCIDENTAL SHOCK HAZARD
,
THIS DEVICE MUST BE PLUGGED INTO A PROPERLY GROUNDED
,
GFI
PROTECTED THREE WIRE RECEPTACLE
AC
CIRCUIT
.
D
O NOT EMPLOY EXTENSION CORDS OF ANY KIND
.
W
HEN THE POWER
CORD IS NOT LONG ENOUGH TO BE SERVICEABLE
,
A LICENSED ELECTRICIAN MUST INSTALL A NEW THREE WIRE GROUNDED
RECEPTACLE IN ACCORDANCE WITH THE
N
ATIONAL
E
LECTRICAL
C
ODE
.
D
O NOT USE A THREE TO TWO PRONG PLUG ADAPTER
WITH THIS DEVICE
.
CAUTION!
D
O NOT USE CHEMICAL CLEANING AGENTS THAT MAY DAMAGE THE MATERIAL USED IN THE
G
RANU
F
LO
I
D
ISSOLUTION
U
NIT
.
A
GENTS WHICH CONTAIN
BENZENE,
TOLUENE,
XYLENE,
ACETONE
OR ANY OTHER
AROMATIC
OR
KETONE
SOLVENTS
MUST BE AVOIDED
.
NOTE:
U
SE
O
NLY
5%
TO
10%
S
ODIUM
H
YPOCHLORITE
(
REGULAR BLEACH
)
TO DISINFECT THE
G
RANU
F
LO
I
D
ISSOLUTION
U
NIT
.
WARNING!
ENSURE
THE
TRANSFER
HOSE
IS
NOT
CONNECTED
TO
A
CONCENTRATE
STORAGE
CONTAINER/TANK.
Summary of Contents for GranuFlo 450385
Page 1: ... GranuFlo CONCENTRATE DISSOLUTION UNIT I Version 2 OPERATOR S MANUAL P N 450385 Rev H ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 38: ...32 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 40: ...34 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 42: ...36 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...38 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 48: ...42 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...P N 450385 Rev H 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 50: ...44 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 450385 Rev H 45 ...
Page 52: ...46 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 450385 Rev H 47 ...
Page 54: ...48 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 450385 Rev H 49 ...
Page 56: ...50 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 58: ...52 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 60: ...54 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 450385 Rev H 55 PAGE INTENTIONALLY LEFT BLANK ...