M A I N T E N A N C E Continued
P/N 450385 Rev. H
25
•
Adjust the output of the Transfer Nozzle to a slow output flow.
•
Place the test strip under the flow for 30 seconds or the time interval recommended by the
manufacture of the test strip being used.
•
Close Transfer Nozzle Valve and press the PAUSE button (the PAUSE indicator light will
illuminate).
•
Ensure that you check the results of the test strip immediately. If the results are higher than 0.1
ppm (RD61: 2006) go to step 9.
•
Before proceeding to step 8, ensure the Drain Hose is secure to prevent the Hose from flailing
around during the DRAIN Operation.
8.
Checking bleach residuals during DRAIN Operation from the Drain Hose.
•
Have chlorine test strip ready for use.
•
Using STEP button step to the DRAIN Operation.
•
Press the RINSE CYCLE side START button.
•
PAUSE Indicator Light will turn off and the Drain Valve will OPEN. Allow water to drain for
30 seconds.
•
Place test strip under the flow for 30 seconds, or the recommended time on the test strip
manufacture’s instruction. Press PAUSE button once you are complete with this step.
•
Ensure that you check the results of the test strip immediately.
9.
If residual bleach levels are higher than ANSI/AAMI Standard limit of <0.1 ppm (RD61: 2006), in any of
the two testing regions, initiate another complete RINSE CYCLE. After the RINSE CYCLE is complete,
start from section 8.3, #5, to check for residual bleach. Continue the RINSE CYCLE and test procedure
until bleach residuals are within ANSI/AAMI Standard limit of <0.1 ppm (RD61:2006) at the end of the
Transfer and Drain Hose.
10.
Once you have attained an acceptable bleach residual reading, ensure the Mixing Tank is empty, turn
OFF power to the GranuFlo I Dissolution Unit, and CLOSE the Main Transfer Ball Valve near the Filter
Housing. Then, connect the Transfer Hose back onto the GranuFlo I Dissolution Unit Transfer Hose
Holder (See Diagram 2: GranuFlo I Dissolution Unit, pg 8).
11.
Remove the Filter Housing and drain all residual water from the Housing. Install a new Filter and
tighten the Filter Housing into place (Figure 21).
Figure 21
12.
Immediately after RINSE CYCLE, make a batch of Dry Acid Product using Section 7, pg 14. Leaving the
GranuFlo I Dissolution Unit with only treated water or wetted with only treated water leaves the Unit
susceptible to bacterial growth.
CAUTION!
D
O NOT ALLOW THE
U
NIT TO REMAIN FULL OF WATER WITHOUT THE ADDITION OF
F
RESENIUS
M
EDICAL
C
ARE
G
RANU
F
LO
®
D
RY
A
CID
P
RODUCT
.
B
ACTERIAL GROWTH MAY OCCUR
.
A. Remove filter housing
B. Insert filter (P/N G84-
202-12)
C. Install filter housing onto
the filter holder.
D. Tighten the filter housing
Summary of Contents for GranuFlo 450385
Page 1: ... GranuFlo CONCENTRATE DISSOLUTION UNIT I Version 2 OPERATOR S MANUAL P N 450385 Rev H ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 38: ...32 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 40: ...34 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 42: ...36 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...38 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 48: ...42 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...P N 450385 Rev H 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 50: ...44 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 450385 Rev H 45 ...
Page 52: ...46 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 450385 Rev H 47 ...
Page 54: ...48 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 450385 Rev H 49 ...
Page 56: ...50 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 58: ...52 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 60: ...54 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 450385 Rev H 55 PAGE INTENTIONALLY LEFT BLANK ...