D I S S O L U T I O N C Y C L E Continued
P/N 450385 Rev H
15
9.
Remove the GranuFlo I Dissolution Unit Small Access
Lid, and slowly add granules as shown in Figure 7. This
minimizes the amount of particulates discharged into
the air while adding granules to the tank. If there is
clumping present in the dry acid, then the larger
Access Lid can be used to load the Tank; but ensure
that the bags are poured slowly into the large access
port to avoid splashing water onto the outside of the
GranuFlo I Dissolution Unit (See Appendix A for further
complete dry acid mixing information, page 43).
Figure 7
10.
After the Dry Acid Concentrate has been added, remove the Large Access lid. Dry off the Final Fill
Sensor of any splash that may have occurred during the Add Granules process (Figure 8).
Figure 8
CAUTION!
B
E CERTAIN THAT THE SAME
C
ATALOG
N
UMBER IS ON ALL CASES
.
CAUTION!
U
SING EYE PROTECTION
,
AND GLOVES IS RECOMMENDED
.
I
F CONTACT WITH EYES
,
RINSE IMMEDIATELY
FOR
15
MINUTES
.
I
F CONTACT WITH SKIN
,
FLUSH WITH PLENTY OF SOAP AND WATER
.
S
EE
M
ATERIAL
S
AFETY
D
ATA
S
HEET
(MSDS)
FOR FURTHER
P
ERSONAL
P
ROTECTIVE
E
QUIPMENT
(PPE)
OR EMERGENCY INSTRUCTIONS FOR
G
RANU
F
LO
®
DRY ACID PRODUCT
.
WARNING!
D
O NOT USE ANY CASE OF DRY ACID CONCENTRATE THAT HAS BEEN OPENED OR TAMPERED WITH
.
I
T
IS IMPORTANT THAT THE ENTIRE CONTENTS OF EACH CASE OF DRY ACID CONCENTRATE IS EMPTIED INTO THE
G
RANU
F
LO
I
D
ISSOLUTION
U
NIT
.
B
E CERTAIN ALL BAGS PER CASE HAVE BEEN ADDED
.
WARNING!
IN ABSENCE OF SIGNIFICANT CLUMPING
,
ONLY USE THE SMALL ACCESS LID WHEN POURING ACID
CONCENTRATE INTO THE MIXER
.
USE OF THE LARGE ACCESS LID WHEN POURING ACID CONCENTRATE MAY RESULT
IN INHALATION INJURY
.
NOTE:
G
RANU
F
LO
®
D
RY
A
CID
P
RODUCT CONTAINS
3
BAGS PER BOX
.
C
ONTENTS OF ALL EIGHT BOXES MUST BE
POURED INTO THE TANK DURING THE
ADD
GRANULES
O
PERATION
.
NOTE:
I
NPUT WATER PRESSURE DIRECTLY AFFECTS HOW FAST
T
HE
G
RANU
F
LO
I
D
ISSOLUTION
U
NIT WILL FILL
.
T
HE MAXIMUM INPUT WATER PRESSURE IS
60
PSI WITH A MINIMUM
½
GALLON PER MINUTE FLOW RATE
.
Locate Final Fill Sensor
Clean Final Fill Sensor with a soft cloth.
Summary of Contents for GranuFlo 450385
Page 1: ... GranuFlo CONCENTRATE DISSOLUTION UNIT I Version 2 OPERATOR S MANUAL P N 450385 Rev H ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 38: ...32 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 40: ...34 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 42: ...36 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...38 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 48: ...42 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...P N 450385 Rev H 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 50: ...44 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 450385 Rev H 45 ...
Page 52: ...46 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 450385 Rev H 47 ...
Page 54: ...48 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 450385 Rev H 49 ...
Page 56: ...50 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 58: ...52 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 60: ...54 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 450385 Rev H 55 PAGE INTENTIONALLY LEFT BLANK ...