D I S S O L U T I O N C Y C L E Continued
16
P/N 450385 Rev. H
11.
Replace the Large and Small Access Lid then press the
Dissolution side START button.
The GranuFlo I Dissolution Unit will proceed to the MIX
Operation.
12.
During the MIX Operation, the solution is mixed for a period of forty-five
(45) minutes allowing the granules to completely dissolve. The GranuFlo I
Dissolution Unit will then automatically step to the DEAERATION Operation.
13.
The DEAERATION
Operation is five (5) minutes during which the entrapped
air (bubbles) are allowed to separate out of the solution. Upon completion,
the GranuFlo I Dissolution Unit will automatically step to FINAL FILL
Operation, and the Final Fill Indicator Light will illuminate.
14.
In the FINAL FILL
Operation, the fill valve will OPEN and the GranuFlo I
Dissolution Unit will fill to the final water level.
15.
When the final water level is reached, the Unit automatically step to
HOMOGENIZE
Operation. Verify that the solution level is touching the Final-
Fill Sensor. CLOSE the water supply valve to the GranuFlo I Dissolution Unit.
16.
During HOMOGENIZE Operation, the Mixer will mix the solution for ten (10)
minutes. When the HOMOGENIZE Operation is complete, the Transfer
Indicator LED will flash. Remove the Large Access Lid and look into the Tank
to make sure the granules have dissolved and the solution is colorless.
17.
Once you have verified the granules are dissolved and the solution is
colorless, the solution is ready for Specific Gravity testing. If the granules
have not dissolved and/or the solution is not colorless, call Technical Service
(1-800-227-2572) for further instructions.
NOTE:
D
O NOT PLACE THE
H
YDROMETER INTO THE
G
RANU
F
LO
I
D
ISSOLUTION
U
NIT AT ANY TIME
.
W
ARNING
:
S
PECIFIC
G
RAVITY TEST MUST BE PERFORMED AFTER EVERY BATCH OF CONCENTRATE THAT IS MADE
.
IN
ADDITION
,
C
ONCENTRATE SAMPLES SHOULD BE TAKEN FROM THE TRANSFER
H
OSE ONLY
.
Summary of Contents for GranuFlo 450385
Page 1: ... GranuFlo CONCENTRATE DISSOLUTION UNIT I Version 2 OPERATOR S MANUAL P N 450385 Rev H ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 38: ...32 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 40: ...34 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 42: ...36 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...38 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 48: ...42 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...P N 450385 Rev H 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 50: ...44 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 450385 Rev H 45 ...
Page 52: ...46 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 450385 Rev H 47 ...
Page 54: ...48 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 450385 Rev H 49 ...
Page 56: ...50 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 58: ...52 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 60: ...54 P N 450385 Rev H PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 450385 Rev H 55 PAGE INTENTIONALLY LEFT BLANK ...