ag V2.0 Technician’s Manual
P/N 490188 Rev. D
General Warnings
Warning:
The concentrate displayed on the screen must match the labels on the
acid container. Make certain there is enough concentrate in the containers to
complete the treatment.
Warning:
The specific concentrate, sodium, and bicarbonate settings must be
prescribed by a physician.
Warning:
Acid and basic bicarbonate hemodialysis concentrate must be diluted
(mixed with purified water as specified in the AAMI standards for water for
dialysis) immediately prior to application only.
Warning:
Use aseptic technique.
Warning:
Always verify the conductivity and approximate pH of the dialysate
solution through independent means before initiating dialysis. Verify that the pH
is normal and that the conductivity is reasonably close to the theoretical value. If it
is not, do not initiate dialysis.
Warning:
Replace a leaking bi
b
ag immediately. Spills can cause damage to
carpeting and other surfaces. To contain such spills, the machine should be on a
spill-tolerant surface. Spills can cause slips and falls; clean up spills immediately.
Caution
: Only the bags manufactured by Fresenius Medical Care may be used in
the bi
b
ag connector.
Note
: When the bi
b
ag connector is installed, the online pressure holding test
becomes mandatory. For more information, see the Online Pressure Holding Test
section of the
2008
®
T Hemodialysis Machine Operator’s Manual
P/N 490122 or
the
2008K@HOME
™
Hemodialysis Machine Operator’s Manual
P/N 490180.