Page 25
bi
b
ag V2.0 Technician’s Manual
P/N 490188 Rev. D
Debug Screens (cont.)
Debug Screen 15 (16) – Bic Mon & Act. Reg
(cont.)
PHT Err
Indicates a bi
b
ag Pressure Holding Test failure.
1 =Yes / 0 = No
AcidPr Cal Er
Indicates an Acid Pressure Calibration Error.
1 =Yes / 0 = No
InletT Cal Er
– (Not Used)
BicT Cal Er
Indicates a Bicarb Temperature Calibration Error.
1 =Yes / 0 = No
V105 Er
Indicates a Valve 105 Error.
1 =Yes / 0 = No
V104 Er
Indicates a Valve 104 or Valve 108 Error.
1 =Yes / 0 = No
Reg O/S Rx
Indicates the bi
b
ag interface board has received the Regulator Pressure service mode
calibration data.
1 = Data Received / 0 = Data not Received.
V105 Er O
Indicates Valve 105 is Stuck Open.
1 =Yes / 0 = No
V105 Er C
Indicates Valve 105 is Stuck Closed.
1 =Yes / 0 = No