ag V2.0 Technician’s Manual
P/N 490188 Rev. D
Debug Screens (cont.)
Debug Screen 14 (15) –
bi
b
ag
(cont.)
Val2 Err
Indicates a bi
b
ag Valve 2 error.
1 =Yes / 0 = No
Cond High
Indicates a bi
b
ag Conductivity High error.
1 =Yes / 0 = No
Cond Low
Indicates a bi
b
ag Conductivity Low error.
1 =Yes / 0 = No
Cond Senr
Indicates a bi
b
ag Conductivity Sensor error. This bit is only set in the Rinse program.
1 =Yes / 0 = No
Temp Senr
Indicates a bi
b
ag Temperature Sensor error. This bit is only set in the Rinse program.
1 =Yes / 0 = No
Bic Lock
Indicates a bi
b
ag Bicarbonate Pump Air Locked error.
1 =Yes / 0 = No
Temperature
Displays the temperature in °C of the sodium bicarbonate concentrate from either the
bi
b
ag or a jug.
Conductivity
Displays the conductivity in mS/cm of the sodium bicarbonate concentrate from either
the bi
b
ag or a jug. This value is temperature compensated.
Pressure
Displays the bi
b
ag pressure in mmHg.
Concentration
Displays the bi
b
ag concentration in g/L during bi
b
ag dialysis.