Page 17
bi
b
ag V2.0 Technician’s Manual
P/N 490188 Rev. D
Debug Screens (cont.)
Debug Screen 14 (15) –
bi
b
ag
(cont.)
Opn Bag
Bag on connector, bi
b
ag door fully open.
1 =Yes / 0 = No
Bag On
Bag on bi
b
ag connector.
1 =Yes / 0 = No
D bd Ver
bi
b
ag interface board (daughter board) software version.
Init State
System is in the bi
b
ag Initial state.
1 =Yes / 0 = No
Tx State
System is in the bi
b
ag Treatment state.
1 =Yes / 0 = No
End State
System is in the bi
b
ag End state.
1 =Yes / 0 = No
Vent
System is running the bi
b
ag vent process.
1 =Yes / 0 = No
DeAirLock
System is running the bi
b
ag deairlock (air lock removal) process.
1 =Yes / 0 = No
Tm Bypass
System is running the bi
b
ag timed bypass process.
1 =Yes / 0 = No
Post Flush
System is running the bi
b
ag post empty flush process.
1 =Yes / 0 = No