Page 23
bi
b
ag V2.0 Technician’s Manual
P/N 490188 Rev. D
Debug Screens (cont.)
Debug Screen 15 – Bic Mon & Act. Reg
Note: Debug Screen 16 on 2008K@HOME
™
Bic Low
This bit is normally set to 0, and is set to 1 when the bicarbonate conductivity (Bic Mon
Cond) has been less than “Bic Lo Th” for 40 balance chamber switches.
Bic Hi
This bit is normally set to 0, and is set to 1 when the bicarbonate conductivity (Bic Mon
Cond) has been greater than “Bic Hi Th” for 40 balance chamber switches.
Bic Cell Cond
The raw uncompensated bicarbonate conductivity being seen at the Bicarbonate
Conductivity Cell 117.
Bic Mon temp
The temperature being seen at the Bicarbonate Temperature Thermistor 116.