Actions
Meanings
Symbol
This alarm appears 10 min. before
battery is fully discharged.
■
Connect the pump to the mains via
the pump holder. Recharge battery
to resume pump operation.
Alarm medium
Minimum battery voltage is not avail-
able.
Battery empty
flashing
Medium priority - Yellow LEDs are flashing
■
End feeding or connect to a filled
feed container.
Feed container is empty.
Empty bag / Air
in line
■
Fill giving set to the end.
▷
See
on page
20.
Air is in the giving set.
■
Open the door and remove dirt
with cloth and soapy water or as
directed by hospital policy (see
on page
36). Allow the pump to dry.
Dirt in sensor area (lower tube guide).
■
Check position of giving set and
insert correctly if necessary.
▷
on page
17.
Giving set not properly connected to
the container.
High priority - Red LEDs are flashing - Alarm sound
■
Note the technical Error code (Err
xyz).
■
To release technical alarms, press
or
for 2 seconds. The
pump will then switch off instantly
(no count-down).
■
Contact your biomedical depart-
ment.
A technical alarm code is displayed
with the “Pump error alarm” drawing.
Technical alarm
45