Feeding stops
Technical alarm
(High priority alarm)
Flashing red LED and
alarm sound (sequences of
10 beeps)
Feeding stops
Fail safe technical alarm
(High priority alarm)
Flashing red LED and
buzzer sound
When a functional alarm or prior information to alarm occurs:
■
to mute alarm sound, press
, see
on page 26;
■
detect the specific problem causing the alarm or prior information to alarm condition, by
looking at the drawing displayed on the pump;
■
to release alarm, press
;
■
make a corrective action (see table below);
■
restart feeding using the
key.
WARNING
Identify displays, symbols and status in the table below, to understand the
meaning and conduct the appropriate action.
8.1.2 Alarm descriptions
Line control
Actions
Meanings
Symbol
Medium priority - Yellow LEDs are flashing
■
Check position of giving set above
and below the pump mechanism
and insert correctly if necessary.
■
Check that the proper set is used
(Amika giving sets only).
▷
on page
17.
Missing giving set or giving set not
properly installed or wrong set in-
stalled.
Giving set
■
Remove dirt with cloth and soapy
water or as directed by hospital
policy.
■
Allow the pump to dry.
▷
on page
38.
Area where Pinch clamp is inserted is
contaminated.