Actions
Meanings
Symbol
The time of message before target
volume is reached can be set in the
menu.
▷
See
Set time for target volume al-
on page 34.
■
End feeding or continue feeding.
Target volume will be reached.
Target volume
almost reached
Medium priority - Yellow LEDs are flashing
■
End feeding or proceed to the next
step.
Alarm
The target volume is reached.
(Complete progress bar)
Target volume
reached
Function control
Actions
Meanings
Symbol
Low priority - Symbol battery is fixed yellow
This message appears 30 min. before
the empty battery alarm.
■
Connect the pump to the mains via
the pump holder. Recharge battery
to resume pump operation.
Minimum battery voltage is not avail-
able.
Battery empty
fixed
Medium priority - Symbol battery is flashing yellow
44