5 Operations
5.1 Power management
5.1.1 Connect to the mains
Before connecting to the mains power supply, check that the mains power supply voltage
corresponds to the value indicated on the label on the back of the pump holder. Do not
exceed the permitted voltage.
WARNING
■
The power outlet must remain accessible at all times to allow emergency
power supply disconnection.
■
The pump and its accessories can only be connected to the mains power
supply with the power cable supplied by Fresenius Kabi.
For connection, described in
on page 21.
5.1.2 Battery operating mode
The device is provided with an internal battery that automatically provides power to the
device in case of power failure or disconnection from the mains power supply. The battery
charges when the pump is connected to mains power supply.
Before starting for the first time, charge the battery for at least 6 hours by plugging in the
power supply cable with the pump powered off.
INFORMATION
During operation, leave the device connected to the power supply to maintain
the battery's charge and maximum capacity, and to maximize battery lifetime
and performance.
The battery icon is always displayed in the status bar. The device can be used while battery
is charging.
A minimum of 24 hours at 25 mL/h and a minimum of 14 hours for flow rate
at 600 mL/h in the following conditions:
■
after battery maximum charge
■
with minimum backlight
■
temperature at 22.5°C ± 2.5°C
■
new battery with maximum 5 cycles of charging and discharging
Battery life
When the pump is connected to the mains (see
on
page 21)
► Battery charges automatically, also during operation
(green)
When the pump is disconnected from the mains (see
on page 22)
► Pump switches to Battery Mode automatically
24